Absorbance (also called optical density) is measured in absorbance units (AU), a dimensionless quantity that reports how much light a sample attenuates at a given wavelength. You enter the intensity of light passing through a blank (I₀) and the intensity passing through your sample (I) — or more commonly, the instrument does it for you — and the absorbance is calculated as A = log₁₀(I₀/I). The result has no physical unit like grams or meters; the “AU” is just a label to remind you it is a logarithmic ratio.
Formula
The fundamental equation is the Beer–Lambert law, which links absorbance to the properties of the sample:
A = ε · c · l
- A = absorbance (absorbance units, AU)
- ε (epsilon) = molar absorptivity (L·mol⁻¹·cm⁻¹) — a property of the molecule at a specific wavelength
- c = concentration of the absorbing species (mol/L)
- l = path length of the light through the sample (cm, typically 1.00 cm in standard cuvettes)
Absorbance is also defined directly from the measured light intensities: A = log₁₀(I₀/I), where I₀ is the light intensity before the sample (measured with a blank) and I is the intensity after the sample. Transmittance, T = I/I₀, is the fraction of light that passes through, and A = −log₁₀(T).
Worked Example
Using the calculator defaults: I₀ = 100 (arbitrary units) and I = 1 (same arbitrary units).
- Calculate transmittance: T = I/I₀ = 1/100 = 0.01, or 1%.
- Calculate absorbance: A = log₁₀(100/1) = log₁₀(100) = 2.000 AU.
- Verify with transmittance: A = −log₁₀(0.01) = 2.000 AU.
The calculator gives A = 2.000 AU and T = 1.00%, matching the manual calculation.
Assumptions
- The light is monochromatic (single wavelength).
- The sample is homogeneous and dilute enough that molecules do not interact.
- The cuvette path length is constant and the windows are clean and parallel.
- The blank corrects for reflections, solvent absorption, and any cuvette effects.
Why Absorbance Has No Ordinary Unit
Many students expect absorbance to be measured in something like “grams per liter” or “percent,” but it is a ratio of two light intensities. Because it is a logarithm, a change of 1 AU means the transmitted light drops by a factor of 10. For example, A = 0 means all light passes through (T = 100%), A = 1 means 10% passes, A = 2 means 1% passes, and A = 3 means 0.1% passes. The “AU” is a convenience label — it is not a dimension like mass or length. In practice, instruments display “Abs” or “AU” on the readout, but the value is simply a number.
How a UV-Vis Spectrophotometer Measures It
A typical UV-visible spectrophotometer has a light source (deuterium lamp for UV, tungsten for visible), a monochromator to select a wavelength, a sample holder, and a detector. The measurement procedure is:
- Zero the instrument with a blank (cuvette filled with solvent only) at your chosen wavelength — this sets I₀.
- Replace the blank with your sample cuvette.
- The detector measures the light intensity reaching it after passing through the sample (I).
- The instrument computes A = log₁₀(I₀/I) and displays the absorbance.
Modern instruments do this automatically, but the principle is the same. If you measure without a blank, you will include the absorbance of the solvent and the cuvette, which is why the blank step is essential.
Choosing the Right Wavelength and Cuvette
For quantitative analysis, you should measure at the wavelength of maximum absorbance (λmax) for your analyte, because that gives the greatest sensitivity and the smallest relative error. To find λmax, run a spectrum (absorbance vs. wavelength) and note the peak. For example, DNA absorbs at 260 nm, proteins at 280 nm, and many colored dyes in the visible range.
Cuvettes matter too: standard glass or plastic cuvettes absorb UV light below about 300 nm, so for UV work you need quartz cuvettes. The path length is usually 1.00 cm, but if you use a different path length, you must account for it in the Beer–Lambert calculation. Always check that your cuvette is clean and free of fingerprints, as these scatter light and inflate absorbance.
Common Problems and Misconceptions
- Absorbance above 2–3 is unreliable. At high absorbance, very little light reaches the detector, and stray light or detector noise dominates. Dilute the sample so A is below ~1.5 for best accuracy.
- Beer–Lambert law fails at high concentration. At high concentrations, molecules interact and the linear relationship between A and c breaks down. Always work in the linear range.
- Using the wrong path length. If you use a 0.5 cm cuvette but assume 1 cm, your calculated concentration will be off by a factor of 2.
- Forgetting to subtract the blank. The blank corrects for solvent and cuvette; without it, your absorbance is too high.
- Confusing absorbance with transmittance. Transmittance is a linear ratio (0–100%), while absorbance is logarithmic (0 to ∞). They are inversely related but not linearly.
Absorbance vs. Optical Density (OD)
In practice, “optical density” (OD) is often used interchangeably with absorbance, especially in microbiology (OD600 for cell cultures) and in some older instruments. Technically, OD refers to the raw instrument reading including scattering and other losses, while absorbance implies a pure absorption measurement. For clear solutions, they are numerically the same. For turbid samples like bacterial cultures, the reading includes light scattering, so it is more correct to call it optical density, not absorbance. The unit is still AU or OD.
Frequently Asked Questions
Is absorbance measured in cm⁻¹?
No. The path length has units of cm, and the molar absorptivity has units of L·mol⁻¹·cm⁻¹, but when you multiply ε × c × l, the units cancel to give a dimensionless number. Absorbance itself has no units; the “AU” is just a label.
What is the difference between absorbance and transmittance?
Transmittance is the fraction of light that passes through the sample (I/I₀), expressed as a percentage (0–100%). Absorbance is the negative logarithm of transmittance: A = −log₁₀(T). So if T = 10%, then A = 1.0 AU.
Why is absorbance called “optical density”?
Historically, the term “optical density” was used in photography and later in spectroscopy to describe how much a sample “denses” or blocks light. It is mathematically identical to absorbance for clear solutions, but in turbid samples it includes scattering, so many prefer “absorbance” for true absorption measurements.
Can absorbance be negative?
Yes, if the sample transmits more light than the blank (e.g., if the blank absorbs more than the sample). This can happen if you forgot to include a component in the blank, or if the sample fluoresces. Negative absorbance usually indicates a measurement error or an improper blank.
What is a good absorbance range for accurate measurements?
Most instruments are most accurate in the range of 0.1 to 1.5 AU. Below 0.1, noise and drift dominate; above 1.5–2.0, stray light and detector non-linearity cause errors. If your sample reads above 2, dilute it and multiply by the dilution factor.
Does the unit of concentration affect the absorbance value?
Yes, because the molar absorptivity ε is defined for concentration in mol/L. If you use a different concentration unit (e.g., mg/mL), you must use a different extinction coefficient (often called absorptivity with units L·g⁻¹·cm⁻¹). Always match the units of ε to the units of c.
Related
Formula used
in f | Incident intensity I₀ | - | 100 | 0 in i | Transmitted intensity I | - | 1 | 0 out a | Absorbance | AU | log10(f / i) | 3 out t | Transmittance | % | (i / f) * 100 | 2
Each line is one input (in, sel, list) or one result (out: label, unit, formula, decimals). How the calculators work
Method: the standard formula for this calculation with the constants and conversion factors written on the page, rounded as each result states. See method and constants, the disclaimer or report an error.
More in this section
- QPP Calculator Canada: How It Works & How to Use It
- PPP Calculator World Bank: How It Works & How to Use It2,000.00
- PPP Calculator: India vs USA Cost Comparison1,000.00 USD
- PPP Calculator USD to INR: How It Works2,335 INR
- PPP Calculator India vs UK: Real Cost Comparison103.63 GBP
- PPP Calculator India vs Dubai: Real Cost Comparison8,954.58 AED