Calculate relative centrifugal force (RCF, or ×g) from a centrifuge rotor’s radius and rotational speed in revolutions per minute (RPM). Enter the distance from the axis of rotation to the sample and the centrifuge speed to get the acceleration relative to standard gravity; check your rotor manual before setting a run.
Formula
The standard practical formula is RCF = 1.118 × 10−5 × r × RPM², where r is the rotational radius in centimetres and RPM is revolutions per minute. The calculator writes the same formula as RCF = 1.118 × r × (RPM / 1000)²; expanding the square gives the coefficient 1.118 × 10−5.
- RCF is relative centrifugal force, expressed as a multiple of standard gravity (×g); it has no physical unit.
- r is the radius in centimetres, measured from the axis of rotation to the sample location specified by your protocol.
- RPM is rotor speed in revolutions per minute.
- Acceleration converts RCF to metres per second squared using standard gravity, g₀ = 9.80665 m/s². Thus, acceleration = RCF × 9.80665.
The constant combines the unit conversions in the circular-motion relationship a = ω²r, with angular speed ω = 2π × RPM / 60 and radius converted from centimetres to metres. RCF is a ratio: an RCF of 1,000 ×g means the sample experiences an acceleration magnitude 1,000 times standard gravity. It does not mean the rotor is turning at 1,000 RPM.
Worked Example
Using the calculator’s default inputs, the radius is 10 cm and the speed is 5,000 RPM.
- Divide the speed by 1,000: 5,000 / 1,000 = 5.
- Square that value: 5² = 25.
- Multiply by the radius and the formula coefficient: 1.118 × 10 × 25 = 279.5.
- The relative centrifugal force is therefore 279.50 ×g, to two decimal places.
- For the acceleration output, multiply by standard gravity: 279.5 × 9.80665 = 2,740.959575 m/s², or 2,740.96 m/s² to two decimal places.
Assumptions
- The rotor turns at a steady speed, and the radius is measured in centimetres from the axis of rotation.
- The reported RCF applies at the radius you enter. A sample occupies a range of radii, so its RCF varies from one point to another.
- Standard gravity is taken as exactly 9.80665 m/s² for converting ×g to acceleration.
- The calculation does not determine safe speed, suitable tubes, run duration, or whether a rotor and centrifuge are compatible.
Step-by-Step
- Find the correct radius. Consult the rotor manual or a validated protocol for the radius definition required. Radius is measured from the centre of rotation (the rotor’s axis) outward to a point in the sample. Depending on the protocol, it may specify the sample’s maximum radius, minimum radius, or a defined average. Do not substitute the rotor’s overall diameter: if you have a diameter measured through the axis, divide it by two to get the radius.
- Convert the radius to centimetres. The formula requires centimetres. Convert millimetres to centimetres by dividing by 10, or inches to centimetres by multiplying by 2.54. For example, 110 mm is 11 cm. Enter the relevant radius value, not a length measured from the rotor’s outer edge unless that is how the specified radius is defined.
- Get the speed in RPM. Use the speed stated by your method or the verified centrifuge setting. If the method gives RCF instead, do not guess an RPM; use the inverse calculation or the centrifuge’s validated conversion for that rotor. Enter RPM as a revolutions-per-minute number, not revolutions per second.
- Enter both values and calculate. Provide the radius in centimetres and speed in RPM. The calculator squares the speed, so speed errors have a large effect: a 10% increase in RPM produces about a 21% increase in RCF at the same radius.
- Read the result with its location in mind. The main answer is RCF in ×g at the radius you entered. Compare it with the target in your protocol and verify that the protocol uses the same radius convention. The secondary result is the equivalent acceleration in m/s².
- Check operating limits before running. Confirm the rotor’s maximum rated speed, the centrifuge’s limits, and the correct rotor, buckets, adapters, and tubes. The calculator only does the arithmetic; it cannot tell whether a chosen speed is safe or appropriate. Follow your lab method, the safety data sheet (SDS) and your supervisor.
Common Mistakes
- Using diameter as radius: this doubles the entered radius and therefore doubles the calculated RCF. Measure from the axis to the specified sample point.
- Mixing units: entering a radius in millimetres as though it were centimetres makes the answer 10 times too high. Convert to centimetres first.
- Confusing RPM and RCF: RPM describes rotational speed; RCF describes acceleration relative to gravity. The same RPM gives different RCF values in rotors with different radii.
- Using the wrong sample location: RCF is not identical throughout a tube. Using the maximum-radius value when a method specifies an average, or vice versa, can cause a mismatch.
- Rounding too early: retain the original radius and speed in the calculation and round the final result for reporting. Small changes in speed matter because RPM is squared.
- Assuming a displayed setting is exact: check whether the instrument or method specifies RPM or RCF, and whether its setting is associated with the rotor currently installed.
Tips
- Record the rotor model or identifier, radius definition, RPM, and calculated RCF in your notes so another person can reproduce the setting.
- When comparing protocols, compare RCF values only after confirming they refer to the same location in the sample.
- If the target is given in ×g, use the rotor’s radius to determine the corresponding RPM. A conversion for one rotor cannot automatically be reused for another.
- Use the rotor manufacturer’s manual or approved documentation for dimensions and operating limits. Never exceed the lowest applicable limit for the rotor, centrifuge, or accessories.
- For a swinging-bucket rotor, use the radius specified by the method for the bucket’s operating position. A radius measured with the bucket hanging at rest may not represent its position during rotation.
Troubleshooting
The result is 10 or 100 times off: check whether the radius was entered in centimetres and whether RPM was entered as revolutions per minute. Check that a diameter was not entered in place of a radius.
Your answer differs from a centrifuge display: verify that the displayed value is RPM or RCF, then check the selected rotor, the radius used by the instrument, and the specific sample position. Instruments and protocols can use different radius conventions, so a difference does not automatically mean the formula is wrong.
You do not know the rotor radius: find it in the rotor documentation or approved method rather than estimating it from a photograph or the outside of the centrifuge. If the specified radius cannot be confirmed, ask the responsible supervisor before using a converted setting.
The speed needed for the target exceeds a rated limit: do not run at that speed. Use only a validated rotor and setting within all applicable limits, or consult the supervisor for an approved alternative. The calculator is not a safety approval.
Frequently Asked Questions
What does RCF mean?
RCF means relative centrifugal force. It expresses acceleration at a point in the rotating sample as a multiple of standard gravity and is commonly written as ×g.
How do I calculate RCF from RPM?
Use RCF = 1.118 × 10−5 × r × RPM², with radius r in centimetres and speed in revolutions per minute.
Can I use the same RPM for every rotor?
No. RCF depends on radius as well as RPM, so different rotor geometries produce different RCF values at the same speed.
Which radius should I enter?
Use the radius definition specified by your protocol or rotor documentation—often the distance from the axis to a defined point in the sample. RCF changes across the sample’s radial depth.
How do I convert RCF back to RPM?
Rearrange the formula: RPM = √[RCF / (1.118 × 10−5 × r)], with RCF as a numerical multiple of ×g and radius in centimetres. Check the resulting speed against all equipment limits.
Does this calculation tell me how long to centrifuge?
No. It calculates force from radius and speed, not run duration or separation performance. Use a validated method for time and other operating conditions.
Related
Formula used
in r | Radius from axis to sample | cm | 10 | 0.001 in rpm | Rotor speed | rpm | 5000 | 0 out rcf | Relative centrifugal force | ×g | 1.118 * r * (rpm / 1000) ^ 2 | 2 out gforce | Acceleration | m/s² | rcf * 9.80665 | 2
Each line is one input (in, sel, list) or one result (out: label, unit, formula, decimals). How the calculators work
Method: the standard formula for this calculation with the constants and conversion factors written on the page, rounded as each result states. See method and constants, the disclaimer or report an error.
More in this section
- MD Calc I:T Ratio Calculator and Explanation0.0500 ratio
- Oxygen Mass Number: Isotope Values Explained16 nucleons
- Mass-to-Light Ratio: Meaning and Calculation1.000 M☉/L☉
- Mass Fraction to Molar Fraction Calculator
- M/z Ratio in Mass Spectrometry: Meaning and Calculation1,000.00 Th
- K2 Molar Mass Calculator and Calculation78.1966 g/mol