molecalculator

Moles & Molar Mass

Relative Centrifugal Force (RCF) Calculator and Formula

Calculator No. 01-2250 · Updated · Formula and assumptions shown below

Relative centrifugal force 280 ×g
RPM for target RCF
15,809 RPM
Result, rounded to a whole number
Rotor radius
10 cm
Set in the form below
Rotational speed
5000 RPM
Set in the form below
You enter
Rotor radius (cm), Rotational speed (RPM), Target RCF (×g)
You get
Relative centrifugal force (×g), RPM for target RCF (RPM)
Rounding
0 dp

For study and lab planning. The result follows from the numbers you enter and the formula shown. Check it against your method before you weigh, mix or dilute anything, follow the safety data sheet and your supervisor, and do not use it for clinical, dosing or medication work.

Relative centrifugal force (RCF), also called the g-force, compares the acceleration produced by a centrifuge rotor with the acceleration due to Earth’s gravity. Enter the rotor radius and rotational speed in revolutions per minute (RPM) to calculate RCF in ×g; the calculator also shows the speed needed for a specified RCF.

Formula

RCF is calculated as RCF = 1.118 × 10−5 × r × RPM2, where r is the distance in centimetres from the rotor’s axis of rotation to the sample, and RPM is the rotor’s rotational speed in revolutions per minute. The calculator uses the algebraically equivalent expression 1.118 × r × (RPM / 1000)2. RCF is a ratio, so it is commonly written as ×g; it does not have a physical unit like metres per second squared.

To find the RPM required for a target RCF, rearrange the formula: RPM = 1000 × √[target RCF / (1.118 × r)]. In the calculator, t is the target RCF in ×g. The radius must be positive, and the target RCF cannot be negative. These equations assume the stated radius and speed describe the same rotor configuration and that the rotor is operating at the entered speed.

Worked Example

Using the default inputs—radius 10 cm, speed 5,000 RPM, and target 2,794 ×g:

  1. Convert the speed for the calculator’s equivalent formula: 5,000 / 1,000 = 5.
  2. Calculate RCF: 1.118 × 10 × 52 = 279.5 ×g.
  3. Calculate the speed for the target: 1,000 × √[2,794 / (1.118 × 10)] ≈ 5,000.45 RPM, which displays as 5,000 RPM when rounded to a whole number.

The first result is the RCF at the entered speed. The second is the approximate speed that produces the entered target RCF at that radius. Because the display rounds to whole numbers, the unrounded value for the second result is slightly above 5,000 RPM.

Assumptions

  • The radius is measured in centimetres from the rotation axis to the relevant point in the sample.
  • The rotor reaches the entered RPM, and the speed is steady during the relevant part of the run.
  • The standard RCF relationship treats the sample’s circular motion around the rotor axis; it does not calculate settling time or separation efficiency.

What RCF Means and Why It Matters

A centrifuge rotates samples around an axis. A sample farther from that axis travels in a larger circle; at a given RPM, its centripetal acceleration is greater. RCF expresses this acceleration as a multiple of standard gravitational acceleration, approximately 9.80665 m/s2. Thus, 1,000 ×g means an acceleration about 1,000 times that reference value—not that the sample’s mass or weight has increased a thousandfold.

RCF is useful because different rotors can generate different forces at the same RPM. A protocol specifying only RPM may not transfer reliably between centrifuges if their rotor radii differ. A force specified as ×g is more comparable, provided the radius convention and location in the sample are understood.

Choosing the Correct Rotor Radius

Rotor radius is not necessarily the rotor’s outside radius or the length of a tube. It is the distance from the axis of rotation to the point where the force is being considered. Depending on the rotor and protocol, a manufacturer may report a radius to the sample bottom, the maximum radius, or the average radius of the sample column. Use the radius definition that matches the protocol’s stated RCF.

For a tube containing liquid, the top and bottom of the sample occupy different radii and therefore experience different RCFs. A stated “maximum RCF” generally refers to the greatest radius reached by the sample; an average RCF uses a representative radius. Do not silently substitute one for the other. Consult the rotor documentation or the method when the radius convention is unclear.

RCF Versus RPM

RPM counts how many complete rotations occur each minute. It describes rotational speed, not the acceleration experienced at a particular radius. RCF combines RPM with radius: doubling RPM makes RCF four times larger, while doubling radius makes it twice as large. This squared dependence is why small speed differences can matter.

  • Same rotor, same radius: a higher RPM produces a higher RCF.
  • Different radii: the same RPM can produce different RCF values.
  • Same target RCF: a rotor with a larger working radius needs a lower RPM than one with a smaller radius, subject to its operating limits.

RCF is sometimes casually called “g-force,” but it should not be confused with gravitational force in a general physics problem. In centrifugation, ×g is a convenient way to state acceleration relative to standard gravity.

Common Errors and Practical Limits

  • Entering millimetres as centimetres: convert the radius first; 100 mm is 10 cm. A tenfold radius error produces a tenfold RCF error.
  • Using the wrong radius point: rotor maximum, average, and sample-bottom radii are not interchangeable.
  • Confusing RPM with RCF: a speed value by itself is not enough to determine RCF; rotor radius is also required.
  • Ignoring speed limits: a calculated RPM is not proof that a rotor, tube, or centrifuge can safely operate at that speed. Check equipment limits and the applicable protocol.

This calculation gives an idealized acceleration based on radius and rotational speed. It does not predict how quickly particles sediment, whether a pellet will form, or how well a separation will work. Those outcomes also depend on factors such as particle size, density, sample viscosity, rotor geometry, run time, and temperature. A formula result should therefore not replace a validated method or equipment instructions.

Frequently Asked Questions

How do I calculate RCF from RPM?
Multiply 1.118 × 10−5 by the rotor radius in centimetres and by RPM squared. For example, 10 cm at 5,000 RPM gives 279.5 ×g.

Can I convert RPM to RCF without knowing the rotor radius?
No. The same RPM produces different RCF values at different radii. Find the radius for the relevant rotor and sample position before converting.

Is RCF the same as g-force?
In centrifuge instructions, RCF and g-force usually refer to acceleration expressed as a multiple of standard gravity. An RCF of 500 ×g means approximately 500 times standard gravitational acceleration.

Why does my centrifuge show a different RCF from this calculator?
Check whether it uses a different rotor radius, a different sample position, or an average rather than maximum RCF. Also confirm the entered speed and radius units.

What radius should I use for a centrifuge tube?
Use the radius specified for the rotor and the RCF convention in your protocol. The sample bottom, sample average, and rotor’s maximum radius may differ, so use the matching documented value rather than guessing.

Related

Formula used
in r | Rotor radius | cm | 10 | 0
in n | Rotational speed | RPM | 5000 | 0
in t | Target RCF | ×g | 2794 | 0
out g | Relative centrifugal force | ×g | 1.118 * r * (n / 1000) ^ 2 | 0
out rpm | RPM for target RCF | RPM | 1000 * sqrt(t / (1.118 * r)) | 0

Each line is one input (in, sel, list) or one result (out: label, unit, formula, decimals). How the calculators work

Method: the standard formula for this calculation with the constants and conversion factors written on the page, rounded as each result states. See method and constants, the disclaimer or report an error.

More in this section

The whole Moles & Molar Mass section →