The RNA purity 260/230 ratio is the absorbance at 260 nm divided by the absorbance at 230 nm, measured in a UV spectrophotometer. For relatively pure RNA, it is often around 2.0–2.2; a lower value can indicate carryover contaminants, but the ratio alone cannot identify a contaminant or prove that RNA is usable.
Formula
Divide the sample’s absorbance at 260 nm by its absorbance at 230 nm:
260/230 ratio = A260 ÷ A230
- A260 is the measured absorbance at 260 nm, in absorbance units (AU).
- A230 is the measured absorbance at 230 nm, also in AU.
- Ratio is a quotient and has no physical unit. Absorbance is conventionally reported without a unit.
The calculator performs this division from the two entered readings. Its result is only as meaningful as the measurements: blanking, sample concentration, instrument performance and the measurement method all matter. If A230 is zero, the ratio is undefined; a very small denominator can also make the result unstable.
Worked Example
Using the calculator’s default readings:
- A260 = 0.22 AU
- A230 = 0.10 AU
- 260/230 ratio = 0.22 ÷ 0.10 = 2.20
The result is 2.20 when rounded to two decimal places, matching the calculator output. It is within the commonly cited range for clean RNA, but it does not establish that the sample is intact, free of all inhibitors, or suitable for a particular experiment.
What the Ratio Indicates
Nucleic acids absorb UV light strongly near 260 nm. Many substances that may be carried through RNA extraction absorb more strongly near 230 nm, so excess absorbance at 230 nm can pull the ratio down. Depending on the extraction chemistry and sample, possible contributors include salts, phenolic compounds, carbohydrates, guanidine-containing reagents, or other organic compounds.
A ratio near 2.0–2.2 is a practical reference range often used for RNA, not a universal pass/fail specification. The expected value can vary with the measurement platform, blank solution, concentration and sample matrix. A reading of 1.7, for example, suggests checking for contamination or measurement problems; it does not tell you which one is responsible. Conversely, a value in the reference range does not rule out contamination that absorbs weakly at these wavelengths.
260/230 Compared with 260/280
The two ratios give complementary clues. The 260/230 ratio is especially useful for spotting absorbance associated with extraction-reagent or salt carryover. The 260/280 ratio compares nucleic-acid absorbance with absorbance near 280 nm, where proteins and some other compounds contribute. Purified RNA commonly has a 260/280 ratio around 2.0, though interpretation also depends on the measurement conditions.
Neither ratio is a direct measure of RNA integrity. A sample can have plausible 260/230 and 260/280 values but still contain degraded RNA. Likewise, a low ratio does not automatically mean the RNA is unusable: the effect depends on the contaminant, its amount and the downstream application. Use the ratio as a screening signal alongside concentration, integrity and experiment-specific checks.
Why Measurements Can Be Misleading
UV absorbance measurements are sensitive to how the sample is prepared and measured. A mismatched blank can add or subtract background absorbance, while a dirty measurement surface, bubbles, evaporation or carryover from a previous sample can distort readings. Very dilute RNA can produce weak signals, making ratios noisy; when A230 is close to the instrument’s measurement limit, small errors in that value can substantially change the quotient.
- Use an appropriate blank that matches the sample’s buffer or solvent.
- Check that both readings are within the instrument’s reliable measurement range.
- Inspect the raw absorbance spectrum, if available, rather than relying only on two reported ratios.
- Repeat a questionable measurement using a clean sample and measurement surface, following the instrument’s instructions.
- Compare results with an independent concentration or integrity method when the sample matters to a sensitive application.
Do not interpret a ratio as a concentration measurement. Concentration is inferred from absorbance using a suitable method and assumptions; this calculator only divides A260 by A230. For bench work, follow your lab method, the safety data sheet (SDS) and your supervisor.
What to Do About a Low Ratio
First confirm that the low value is reproducible and not caused by an unsuitable blank, weak signal or measurement artifact. Then consider the extraction method and the likely carryover. If a downstream assay is sensitive to inhibitors, consult its requirements and your laboratory’s RNA cleanup or re-extraction guidance. Cleanup may remove contaminants, but it can also reduce RNA yield; do not assume that a ratio improvement guarantees better integrity or performance.
Choose follow-up checks according to the intended use. Fluorescence-based quantification can provide a more selective estimate of nucleic-acid concentration than UV absorbance in some situations, but it does not by itself identify every contaminant. An electrophoretic or other validated integrity assessment addresses a different question: whether RNA is degraded. No single metric answers all three questions of amount, purity and integrity.
Frequently Asked Questions
What is a good 260/230 ratio for RNA?
A value around 2.0–2.2 is commonly cited for relatively pure RNA. Treat it as a guideline rather than a universal cutoff, and account for the measurement method and downstream assay.
What does a low RNA 260/230 ratio mean?
It means absorbance at 230 nm is high relative to absorbance at 260 nm. This can be consistent with reagent, salt or other carryover, but the ratio cannot identify the cause on its own.
Can RNA be used if its 260/230 ratio is low?
Sometimes. Suitability depends on the likely contaminant, its effect on the intended assay, RNA integrity and the assay’s tolerance. Check the application’s specifications and consider appropriate follow-up testing.
Does a good 260/230 ratio mean RNA is intact?
No. Purity ratios assess relative absorbance, not RNA fragmentation. Use an integrity assessment suited to the sample and experiment.
Why is my 260/230 ratio unusually high?
A high result can occur when A230 is very small, noisy or affected by blanking or measurement error. Review the raw readings and spectrum, confirm the blank and repeat the measurement if needed.
Limits of the Calculator
This calculator reports only A260 ÷ A230. It does not correct the readings, estimate RNA concentration, diagnose a contaminant, assess degradation or determine whether a sample meets a specific assay’s acceptance criteria. Enter absorbance readings from the same sample and measurement, and interpret the result with the method’s quality controls and application requirements.
Related
Formula used
in a260 | Absorbance at 260 nm | AU | 0.22 | 0 in a230 | Absorbance at 230 nm | AU | 0.1 | 0 out r | RNA purity 260/230 ratio | ratio | a260 / a230 | 2
Each line is one input (in, sel, list) or one result (out: label, unit, formula, decimals). How the calculators work
Method: the standard formula for this calculation with the constants and conversion factors written on the page, rounded as each result states. See method and constants, the disclaimer or report an error.
More in this section
- Isotope Percent Abundance Calculator and How to Use It75.91 %
- Percentage Abundance Calculator and How to Find It
- Isotope Fractional Abundance Calculator and Formula0.2409 fraction
- Percent Abundance Formula: Calculate Isotope Percentages25.00 %
- Isotope Percent Abundance Calculator and How to Use It50.00 %
- Isotope Percent Abundance Calculator and How-To Guide75.91 %