RNA Purity Absorbance Ratios Explained

Calculator No. 01-3580 · Updated · Formula and assumptions shown below

A260/A280 purity ratio 2.00 ratio
A260/A230 purity ratio
2.20 ratio
= Absorbance at 260 nm ÷ Absorbance at 230 nm
Absorbance at 260 nm
0.2 AU
Set in the form below
Absorbance at 280 nm
0.1 AU
Set in the form below
You enter
Absorbance at 260 nm (AU), Absorbance at 280 nm (AU), Absorbance at 230 nm (AU)
You get
A260/A280 purity ratio (ratio), A260/A230 purity ratio (ratio)
Rounding
2 dp

For study and lab planning. The result follows from the numbers you enter and the formula shown. Check it against your method before you weigh, mix or dilute anything, follow the safety data sheet and your supervisor, and do not use it for clinical, dosing or medication work.

RNA purity is commonly assessed by the absorbance ratios A260/A280 and A260/A230. For relatively clean RNA, A260/A280 is typically about 2.0 and A260/A230 is often about 2.0–2.2, but these are screening guides—not proof of integrity or fitness for a particular experiment.

Formula

The calculator divides the absorbance measured at 260 nm by the absorbance measured at either 280 nm or 230 nm:

  • A260/A280 = A260 ÷ A280
  • A260/A230 = A260 ÷ A230

A260, A280, and A230 are absorbance readings at 260, 280, and 230 nanometres, respectively. Absorbance is dimensionless (often displayed as AU), so each ratio is also dimensionless. The readings should come from the same sample and measurement setup, with an appropriate blank. The calculator assumes positive denominator readings; a zero reading cannot be used as a divisor.

Worked Example

Using the default readings in the calculator:

  • A260/A280 = 0.200 ÷ 0.100 = 2.00
  • A260/A230 = 0.200 ÷ 0.091 = 2.1978…, displayed as 2.20

These values fall near commonly cited expectations for relatively pure RNA. The A260/A230 result is a little above 2.2 before rounding; treat it as a screening result, not an exact pass/fail verdict.

Assumptions

  • All three readings represent the same RNA sample and use the same blanking and instrument conditions.
  • Absorbance values are positive, and the measured sample is within the instrument’s reliable range.
  • Reference ranges are approximate and depend on measurement conditions and downstream use.

What the Ratios Indicate

Nucleic acids absorb ultraviolet light strongly near 260 nm. Proteins and some other contaminants absorb more strongly near 280 nm, while several common extraction reagents and residual salts can contribute absorbance around 230 nm. Comparing readings at these wavelengths gives a quick indication of whether the sample’s spectrum resembles that expected for nucleic acid.

For purified RNA, an A260/A280 ratio close to 2.0 is commonly expected. A lower value can be consistent with protein, phenol, or other absorbing contaminants, although measurement error and sample conditions can also affect it. A260/A230 is often expected to be around 2.0–2.2. A low value may point to carryover of substances such as guanidinium salts, phenol, carbohydrates, or other compounds used during extraction. No single ratio identifies a contaminant with certainty.

Ratios are relative measurements, not concentration measurements. A sample can have an apparently good ratio but contain very little RNA. Conversely, a concentrated or dilute sample can give less reliable ratios if its absorbance is outside the method’s useful range. Use a suitable concentration measurement as well when RNA quantity matters.

How to Use the Readings

Enter the absorbance values reported for the sample at 260, 280, and 230 nm. If an instrument reports ratios directly, compare them with the calculated values as a basic consistency check. Keep the same sample path and blanking approach across readings; do not combine values from different samples or use a concentration value in place of absorbance.

Use the ratios as one part of a quality assessment:

  1. Check that the blank and sample measurement were appropriate and that the sample was within the instrument’s reliable range.
  2. Review A260/A280 and A260/A230 together rather than relying on one value.
  3. Check RNA concentration separately, using a method suited to the sample and the required accuracy.
  4. Assess RNA integrity with an appropriate electrophoretic or other validated integrity method when degradation would affect the experiment.
  5. Consider the requirements of the downstream application; a ratio that is acceptable for one task may not be sufficient for another.

For bench work, follow your lab method, the safety data sheet (SDS) and your supervisor.

Interpreting Low or Unusual Ratios

A low A260/A280 ratio can suggest protein or phenol contamination, but it does not establish the cause. A low A260/A230 can be associated with residual extraction chemicals, salts, or other absorbing material. If either ratio is unexpected, first check for practical causes: an unsuitable blank, a dirty measurement surface, a sample that is too dilute, carryover from the extraction, or a reading near the instrument’s detection limit.

Do not automatically discard a sample or repeat purification solely because a ratio differs slightly from a reference value. Re-measurement and an independent check can help distinguish a true sample issue from a measurement problem. Additional cleanup may improve contaminant carryover but can also reduce RNA recovery; the best choice depends on the sample and the downstream assay. Avoid treating a single threshold as universal.

Ratios Do Not Measure RNA Integrity

A common misconception is that a ratio near 2.0 means the RNA is intact. Absorbance ratios assess spectral purity, not the size or degradation state of RNA molecules. RNA that has fragmented can still produce plausible A260/A280 and A260/A230 values. To evaluate integrity, use a method that examines RNA size distribution or an assay-specific integrity metric.

Likewise, these ratios do not prove that a sample is free of DNA. DNA also absorbs near 260 nm, so a sample containing DNA can still have an apparently reasonable absorbance ratio. If DNA contamination matters, use an appropriate DNA-sensitive check or a validated removal and verification strategy.

Frequently Asked Questions

What is a good A260/A280 ratio for RNA?
A value near 2.0 is a commonly used reference for relatively pure RNA. It is an approximate guide, not a universal acceptance limit.

What is a good A260/A230 ratio for RNA?
Values around 2.0–2.2 are often expected for clean RNA. A lower value can indicate carryover contaminants, but the ratio alone cannot identify which substance is present.

Can RNA be good if its A260/A280 ratio is low?
Possibly. A low result may reflect contamination, but dilution, blanking, or measurement limitations can also affect it. Interpret the reading with concentration, sample history, and downstream requirements.

Do absorbance ratios show whether RNA is degraded?
No. They indicate spectral relationships, not RNA fragment size. Use an integrity assessment when degradation is important to the experiment.

Can a good ratio rule out DNA contamination?
No. DNA absorbs at a similar wavelength to RNA and may not be distinguished by these two ratios. A DNA-specific check may be needed.

Related

Formula used
in a260 | Absorbance at 260 nm | AU | 0.200 | 0
in a280 | Absorbance at 280 nm | AU | 0.100 | 0
in a230 | Absorbance at 230 nm | AU | 0.091 | 0
out r28 | A260/A280 purity ratio | ratio | a260 / a280 | 2
out r23 | A260/A230 purity ratio | ratio | a260 / a230 | 2

Each line is one input (in, sel, list) or one result (out: label, unit, formula, decimals). How the calculators work

Method: the standard formula for this calculation with the constants and conversion factors written on the page, rounded as each result states. See method and constants, the disclaimer or report an error.