Specific UV absorbance (SUV) is the absorbance of a solution at a given wavelength divided by the concentration of the absorbing substance, usually expressed in L·g⁻¹·cm⁻¹. You enter the measured absorbance, the sample concentration, and the path length of the cuvette, and the calculator returns the specific absorbance coefficient. This value is a property of the molecule itself, independent of concentration and path length, making it useful for identifying compounds and checking purity.
Formula
The specific UV absorbance is derived from the Beer–Lambert law:
A = ε · c · l
where:
- A = absorbance (dimensionless, often called AU or OD)
- ε = specific absorbance coefficient (L·g⁻¹·cm⁻¹) or molar absorptivity (L·mol⁻¹·cm⁻¹)
- c = concentration (g/L or mol/L)
- l = path length of the light through the sample (cm)
Rearranged for the specific absorbance:
ε = A / (c · l)
In the calculator above, the first output is the specific absorbance in L·g⁻¹·cm⁻¹, using concentration in g/L. If you also enter the molar mass, the second output converts that to molar absorptivity (L·mol⁻¹·cm⁻¹) by multiplying by the molar mass.
Worked Example
Suppose a protein solution has an absorbance of 0.850 at 280 nm, a concentration of 0.050 g/L, and is measured in a 1.0 cm cuvette.
Using the formula:
ε = A / (c · l) = 0.850 / (0.050 × 1.0) = 0.850 / 0.050 = 17.0 L·g⁻¹·cm⁻¹
If the protein has a molar mass of 50,000 g/mol, the molar absorptivity is:
ε_molar = 17.0 × 50,000 = 8.5 × 10⁵ L·mol⁻¹·cm⁻¹
This matches the calculator output when you enter the default values.
Assumptions
- The solution is dilute enough that the Beer–Lambert law holds (no aggregation or concentration-dependent effects).
- The absorbance is measured at a single wavelength with a monochromatic light source.
- The path length is accurately known (standard cuvettes are 1.0 cm).
- The solvent and buffer do not absorb significantly at the measurement wavelength.
- The concentration is known independently (e.g., from dry weight or a standard curve).
When to Use Specific UV Absorbance
Specific UV absorbance is most often used for:
- Protein quantification – at 280 nm, aromatic residues (tryptophan, tyrosine) absorb UV light. The specific absorbance (often called the extinction coefficient) is used to convert absorbance to concentration.
- Purity assessment – the ratio of absorbance at 260 nm to 280 nm (A260/A280) is a common check for nucleic acid contamination in protein samples, and vice versa.
- Characterizing nucleic acids – DNA and RNA have specific absorbance values at 260 nm, often expressed as an extinction coefficient of about 0.020 (µg/mL)⁻¹ cm⁻¹ for double-stranded DNA.
- Quality control – verifying that a batch of a compound has the same absorbance properties as a reference standard.
In each case, the specific absorbance is a constant for a given molecule under defined conditions (wavelength, pH, ionic strength, temperature).
How to Read the Result
The result is the absorbance you would expect for a 1 g/L solution in a 1 cm path length cell. A higher value means the molecule absorbs more strongly at that wavelength. For example:
- Most proteins have specific absorbance at 280 nm between 0.5 and 2.0 L·g⁻¹·cm⁻¹ (depending on tryptophan content).
- Pure double-stranded DNA has a specific absorbance of about 20 L·g⁻¹·cm⁻¹ at 260 nm (or 0.020 per µg/mL).
- Small organic molecules vary widely – some have values below 1, others above 100.
If you know the molar mass, the molar absorptivity (ε) is more commonly used in chemistry. It is the absorbance of a 1 mol/L solution in a 1 cm cell.
Common Mistakes and Pitfalls
- Unit mix-ups – concentration in mg/mL instead of g/L gives a result 1000 times too large. Always convert to g/L (1 mg/mL = 1 g/L).
- Path length – some microcuvettes have a 0.2 cm or 0.5 cm path, not 1.0 cm. Check the cuvette specifications.
- Wavelength – the specific absorbance is only valid at the exact wavelength stated. Using a different wavelength (e.g., 280 vs 275 nm) changes the value.
- Solvent baseline – always zero the spectrophotometer with a blank containing the same buffer and cuvette. A dirty cuvette or turbid sample adds absorbance.
- Concentration errors – if the concentration is wrong (e.g., due to inaccurate weighing or dilution), the specific absorbance will be wrong even though the formula is correct.
Limitations of the Formula
The Beer–Lambert law is linear only at low to moderate absorbance (typically below 1.5–2.0 AU). Above that, stray light and detector nonlinearity cause deviations. Also, the specific absorbance depends on the solvent, pH, and temperature. For proteins, the extinction coefficient at 280 nm is often calculated from the amino acid sequence (using the Edelhoch method) rather than measured directly, because pure dry protein is hard to obtain. Always follow your lab method, the safety data sheet (SDS) and your supervisor when performing actual measurements.
Frequently Asked Questions
What is the difference between specific absorbance and molar absorptivity?
Specific absorbance uses concentration in g/L, so the unit is L·g⁻¹·cm⁻¹. Molar absorptivity uses concentration in mol/L, with units L·mol⁻¹·cm⁻¹. To convert, multiply the specific absorbance by the molar mass (g/mol). The two are both constants for a molecule, but molar absorptivity is more common in chemistry because it directly relates to molecular structure.
Why is the A260/A280 ratio used for protein purity?
Nucleic acids absorb strongly at 260 nm, while proteins absorb mainly at 280 nm. A pure protein sample typically has an A260/A280 ratio of about 0.5–0.6. If the ratio is higher (e.g., >0.8), it suggests nucleic acid contamination. Conversely, pure DNA has an A260/A280 ratio of about 1.8–2.0. This is a quick check, not a quantitative measurement.
Can I use specific absorbance to measure concentration directly?
Yes, if you know the specific absorbance (ε) for your molecule at the chosen wavelength and path length, you can rearrange the formula to c = A / (ε · l). This is the basis of UV spectrophotometry. However, you need an independently determined ε – either from a standard, the literature, or a calculated value.
Does the specific absorbance change with pH or buffer?
Yes, for molecules with ionizable groups that absorb UV light (e.g., tyrosine, tryptophan, phenolic compounds), the absorbance can change with pH because protonation alters the electronic structure. Always measure and report the specific absorbance under defined buffer conditions (pH, ionic strength, temperature).
What is a typical specific absorbance for DNA at 260 nm?
Double-stranded DNA has a specific absorbance of about 20 L·g⁻¹·cm⁻¹ at 260 nm (equivalently, 0.020 per µg/mL in a 1 cm cell). Single-stranded RNA is about 25 L·g⁻¹·cm⁻¹. These values are for neutral pH and low salt; they can vary slightly with sequence and buffer.
Why does my measured absorbance exceed 2.0 and become unreliable?
At high absorbance, the spectrophotometer may not have enough light reaching the detector, leading to nonlinear response and stray light errors. Dilute the sample so the absorbance falls between 0.1 and 1.5 AU for the most accurate readings. If you cannot dilute, use a shorter path length cuvette.
Related
Formula used
in a | Absorbance (AU) | - | 0.850 | 0 | 4 in c | Concentration | g/L | 0.050 | 0 | 1000 in l | Path length | cm | 1.0 | 0.1 | 10 out suv | Specific UV absorbance | L/(g·cm) | a / (c * l) | 4 out e | Molar absorptivity (if MW known) | L/(mol·cm) | a / (c * l) | 4
Each line is one input (in, sel, list) or one result (out: label, unit, formula, decimals). How the calculators work
Method: the standard formula for this calculation with the constants and conversion factors written on the page, rounded as each result states. See method and constants, the disclaimer or report an error.
More in this section
- ug to ppb Calculator: Convert Micrograms per Liter1.00 ppb
- Grams to Moles Chemistry: How to Convert0.2000 mol/L
- Convert ug/l to ppb Calculator
- CPP Calculator Ontario: How CPP Is Calculated162.12 $/month
- Grams to Moles to Molecules: How to Convert0.2000 mol/L
- Organic Chemistry Reaction Calculator: How It Works