This page explains the 260/280 ratio, the standard spectrophotometric check for the purity of nucleic acid samples. You enter the absorbance readings at 260 nm and 280 nm, and the calculator returns the ratio that tells you whether a DNA or RNA preparation is contaminated by protein or other absorbing compounds.
What the 260/280 Ratio Actually Measures
The 260/280 ratio is a dimensionless number obtained by dividing the absorbance of a sample at 260 nm by its absorbance at 280 nm. Nucleic acids (DNA and RNA) absorb ultraviolet light most strongly at 260 nm because of the aromatic rings in their nitrogenous bases. Proteins, in contrast, absorb most strongly at 280 nm because of the aromatic amino acids tryptophan and tyrosine. A pure nucleic acid preparation therefore has a high 260/280 ratio, while a preparation contaminated with protein has a lower ratio.
The ratio is not a direct measure of concentration. It is a purity index. The most commonly cited reference values are ~1.8 for pure double-stranded DNA and ~2.0 for pure RNA. These numbers are empirical, not derived from a fundamental law. They depend on the buffer, the pH, the instrument, and even the sequence of the nucleic acid, so they should be treated as guidelines rather than absolute thresholds.
History and Why 260 and 280 nm Were Chosen
The choice of 260 nm and 280 nm dates to the mid-20th century, when UV spectrophotometry became a routine laboratory tool. Researchers observed that nucleic acids have a strong absorbance peak near 260 nm, while proteins have a peak near 280 nm. The ratio was popularized as a quick, inexpensive way to assess purity without needing to run a gel or an assay. The technique was formalized in standard biochemistry protocols and remains a default quality-control step in molecular biology labs today.
It is important to note that the ratio is not a measure of the amount of DNA or RNA. It only tells you whether the sample is relatively free of contaminants that absorb at 280 nm. A sample with a ratio of 1.8 could still be degraded, or contain salts, or have other issues that the 260/280 check does not detect.
How to Use the Calculator and Read the Result
To use the calculator, enter the absorbance values you obtained from your spectrophotometer. The instrument will typically give you A260 and A280 directly. If you have a dilution factor, you should apply it to the concentration calculation separately, but the ratio is unaffected by dilution because both absorbances are scaled equally.
Interpret the result as follows:
- ~1.8 for DNA: Generally considered pure double-stranded DNA.
- ~2.0 for RNA: Generally considered pure RNA.
- Significantly lower than 1.8: Protein contamination is likely, or the sample may contain phenol or other organic compounds that absorb at 280 nm.
- Higher than 2.0: This can indicate RNA contamination in a DNA sample, or the presence of chaotropic salts, or simply that the baseline was not properly zeroed.
Remember that the ratio is only meaningful if the absorbance readings are within the linear range of the instrument, typically between 0.1 and 1.5 absorbance units. If your A260 is above 2, dilute the sample and re-read.
Formula and Worked Example
The formula is simply:
260/280 ratio = A260 / A280
where A260 is the absorbance at 260 nm and A280 is the absorbance at 280 nm. Both are unitless.
Using the default values from the calculator (A260 = 1.2, A280 = 0.65):
Ratio = 1.2 / 0.65 = 1.846… ≈ 1.85
The calculator displays 1.85, which is close to the ideal for DNA. If this were an RNA sample, you would expect a value near 2.0, so a result of 1.85 would suggest slight protein contamination or a buffer effect.
Assumptions and Limitations
The 260/280 ratio is based on several assumptions that are often not stated:
- The sample is measured in a neutral pH buffer (pH ~7). At acidic pH, the ratio can drop by 0.3 or more.
- The spectrophotometer is properly blanked with the same buffer used to dilute the sample.
- The absorbance is measured in a quartz cuvette, since plastic cuvettes absorb in the UV range.
- The nucleic acid is in solution, not bound to a membrane or in a gel.
The ratio does not detect all contaminants. For example, guanidine salts (often used in RNA extraction) absorb at 260 nm and can inflate the ratio. Carbohydrates and some residual reagents do not absorb strongly at either wavelength, so they can go undetected. For a more comprehensive purity assessment, many labs also check the 260/230 ratio (expected ~2.0–2.2) and run an agarose gel to check integrity.
Common Problems and Misconceptions
One common misconception is that a 260/280 ratio of exactly 1.8 guarantees pure DNA. In reality, a degraded DNA sample can still give a 1.8 ratio. The ratio only tells you about contamination that absorbs at 280 nm; it says nothing about the size or integrity of the nucleic acid.
Another issue is that the ratio can be affected by the pH of the buffer. TE buffer (Tris-EDTA) has a pH around 8, which can slightly lower the ratio compared to water. If you measure in water, the pH may be acidic due to dissolved CO2, which can also lower the ratio. Always measure in the same buffer that you use for your downstream application, and always blank with that buffer.
Finally, many people forget that the absorbance values must be within the linear range. If the sample is too concentrated, the detector may be saturated and the ratio will be unreliable. Dilute the sample so that A260 is between 0.1 and 1.0 for the most accurate reading.
Frequently Asked Questions
What is a good 260/280 ratio for DNA?
A ratio of 1.8 is considered pure for double-stranded DNA. Values between 1.7 and 1.9 are generally acceptable for most downstream applications. If your ratio is below 1.7, you may need to re-purify the sample.
What does a 260/280 ratio of 2.0 mean?
For RNA, a ratio of 2.0 is considered pure. For DNA, a ratio of 2.0 may indicate RNA contamination, but it could also be due to the buffer or the instrument. Check the 260/230 ratio and run a gel to confirm.
Why is my 260/280 ratio low?
A low ratio (below 1.7 for DNA) usually indicates protein contamination. It can also be caused by phenol, which absorbs at 280 nm, or by measuring in an acidic buffer. Re-purify the sample and re-measure in a neutral buffer.
Can I use the 260/280 ratio to measure concentration?
No. The ratio is only a purity indicator. To measure concentration, you use the A260 value with the extinction coefficient: for dsDNA, 1 absorbance unit at 260 nm corresponds to 50 ng/µL; for RNA, it is 40 ng/µL; for single-stranded DNA, it is 33 ng/µL.
Does the 260/280 ratio work for all nucleic acids?
It works for DNA and RNA, but the expected values differ. For oligonucleotides, the ratio can vary significantly depending on the base composition, so it is less reliable. For those, use a different method such as a fluorescent dye assay.
Why is my 260/280 ratio higher than 2.0?
A ratio above 2.0 can be caused by RNA contamination in a DNA sample, or by the presence of chaotropic salts like guanidine, which absorb at 260 nm. It can also happen if the blank is incorrect or the sample is too dilute. Check the 260/230 ratio and re-purify if necessary.
Related
Formula used
in a260 | Absorbance at 260 nm | - | 1.2 | 0 in a280 | Absorbance at 280 nm | - | 0.65 | 0 out ratio | 260/280 ratio | - | a260 / a280 | 2
Each line is one input (in, sel, list) or one result (out: label, unit, formula, decimals). How the calculators work
Method: the standard formula for this calculation with the constants and conversion factors written on the page, rounded as each result states. See method and constants, the disclaimer or report an error.
More in this section
- QPP Calculator Canada: How It Works & How to Use It
- PPP Calculator World Bank: How It Works & How to Use It2,000.00
- PPP Calculator: India vs USA Cost Comparison1,000.00 USD
- PPP Calculator USD to INR: How It Works2,335 INR
- PPP Calculator India vs UK: Real Cost Comparison103.63 GBP
- PPP Calculator India vs Dubai: Real Cost Comparison8,954.58 AED