To calculate protein concentration from A280 (absorbance at 280 nm), use the Beer-Lambert law: concentration (mg/mL) = A280 / (ε × path length), where ε is the protein’s extinction coefficient in (mg/mL)−1 cm−1. Alternatively, if you have the molar extinction coefficient (M−1 cm−1), the molar concentration = A280 / (ε_molar × path length). This page provides a calculator that performs both calculations based on your input values.
Formula
The Beer-Lambert law relates absorbance to concentration:
A = ε × c × l
- A = absorbance at 280 nm (unitless, typically 0.1–1.0 for accurate readings)
- ε = molar extinction coefficient (M−1 cm−1), a constant specific to each protein
- c = molar concentration (mol/L)
- l = path length of the cuvette (cm, usually 1 cm)
Rearranged: c = A / (ε × l). To get mass concentration (mg/mL), multiply by the molecular weight (g/mol) and divide by 1000 (since 1 mol/L = MW g/L = MW/1000 mg/mL).
Worked Example
Using the default values from the calculator above:
- A280 = 0.5
- ε = 25,000 M−1 cm−1
- l = 1 cm
- MW = 50,000 g/mol
Step 1: Calculate molar concentration:
c = 0.5 / (25,000 × 1) = 0.5 / 25,000 = 2.0 × 10−5 mol/L
Step 2: Convert to mg/mL:
mg/mL = (2.0 × 10−5 mol/L) × 50,000 g/mol × (1 L / 1000 mL) = 1.0 mg/mL
So the protein concentration is 1.0 mg/mL, which matches the calculator’s output (rounded to 4 decimals).
Assumptions
- The protein solution is pure (no other UV-absorbing contaminants like DNA or phenol).
- The absorbance is measured in a 1 cm path length cuvette at 280 nm.
- The extinction coefficient is known for the specific protein under your buffer conditions.
- The solution is dilute enough that Beer-Lambert law holds (A < 1.5 typically).
Background: Why 280 nm?
Proteins absorb UV light at 280 nm primarily due to the aromatic amino acids tryptophan (Trp) and tyrosine (Tyr), and to a lesser extent phenylalanine (Phe) and disulfide bonds. The absorbance is proportional to the number of these residues, so each protein has a unique extinction coefficient. This method is label-free, non-destructive, and requires only a UV spectrophotometer, making it a standard first step in protein quantification.
How to Choose the Right Extinction Coefficient
You have three options:
- Sequence-based prediction: Use tools like ProtParam or ExPASy to compute ε from the amino acid composition (e.g., ε = (number of Trp × 5,500) + (number of Tyr × 1,490) + (number of Cys × 125) M−1 cm−1). This is a good estimate for denatured proteins.
- Experimental determination: If you have a purified protein, measure A280 of a solution with known concentration (e.g., by amino acid analysis or dry weight) to derive ε.
- Literature values: Many proteins have published ε values (e.g., BSA has ε = 43,824 M−1 cm−1 at 280 nm).
When in doubt, use the sequence-based method, but note that buffer composition (pH, ionic strength) can slightly affect ε.
Common Problems and How to Avoid Them
- Contaminants absorbing at 280 nm: Nucleic acids absorb strongly at 260 nm but also have some absorbance at 280 nm. The A260/A280 ratio for pure protein is ~0.57; if higher, DNA may be present. Use a correction factor or purify further.
- Non-linear absorbance: If A280 > 1.5, the signal may be saturated. Dilute the sample so A280 is between 0.1 and 1.0.
- Wrong path length: Microcuvettes or plate readers may have path lengths other than 1 cm. Always check and adjust l.
- Buffer blanks: Always zero the spectrophotometer with the same buffer used to dissolve the protein, to subtract background absorbance.
- Protein precipitation or aggregation: This can scatter light, causing apparent absorbance. Centrifuge or filter the sample before measurement.
Alternative Methods: A205, Bradford, BCA
A280 is convenient but not always accurate. For dilute samples or those with few aromatic residues, consider:
- A205: Measures peptide bonds at 205 nm, more sensitive but requires a special cuvette (quartz) and is affected by buffer components.
- Bradford assay: Dye-binding method, compatible with many buffers, but varies with protein composition.
- BCA assay: Copper-based, more tolerant to detergents, but not compatible with reducing agents.
Each method has trade-offs; A280 is best for purified proteins with known ε.
Cost and Equipment Considerations
A280 measurement requires a UV spectrophotometer capable of reading at 280 nm, which is standard in most labs. Quartz cuvettes are needed for UV (plastic absorbs UV). The cost is minimal beyond the instrument. The main cost is the time to determine the extinction coefficient if it’s not known.
Safety and Best Practices
Always wear appropriate PPE (gloves, lab coat, safety glasses) when handling protein solutions, especially if they are of biological origin. Follow your lab’s safety data sheets (SDS) for all reagents. For UV light, avoid direct exposure to eyes and skin. Ensure the spectrophotometer is calibrated according to the manufacturer’s instructions.
Frequently Asked Questions
What is a typical extinction coefficient for a protein?
It varies widely. For a protein with average tryptophan and tyrosine content, ε is often between 20,000 and 50,000 M−1 cm−1. For example, BSA has ε ≈ 43,824 M−1 cm−1. Always use the specific value for your protein, not a generic one.
Can I use A280 to measure protein concentration in a cell lysate?
Not reliably, because lysates contain DNA, RNA, and other UV-absorbing molecules. A280 works best for purified proteins. For crude samples, use a colorimetric assay like Bradford or BCA.
How do I correct for nucleic acid contamination?
Measure A260 and A280. The ratio A260/A280 for pure protein is ~0.57, for pure DNA is ~1.8. If your ratio is higher, you can use a correction formula: protein concentration (mg/mL) = 1.55 × A280 − 0.76 × A260 (a common empirical correction for BSA). However, this is approximate; better to remove the contaminant.
What if my sample has a detergent that absorbs at 280 nm?
Some detergents (e.g., Triton X-100) absorb UV. Use a buffer blank containing the same detergent concentration, or switch to a method less affected by detergents, like Bradford (though some detergents interfere too).
Does pH affect the extinction coefficient?
Yes, pH can alter the ionization state of tyrosine residues, which shifts absorbance. For accurate measurements, use a consistent buffer and pH. If you change pH, re-determine ε.
Can I use this calculator for protein solutions in 6 M guanidine-HCl?
Yes, but the extinction coefficient changes in denaturing conditions. Use the ε calculated for denatured protein (based on sequence, ignoring disulfide bonds) and ensure the buffer blank contains the same denaturant.
Related
Formula used
in a | Absorbance at 280 nm (A280) | - | 0.5 | 0 | 4 in e | Molar extinction coefficient (ε) | M^-1 cm^-1 | 25000 | 0 | 1000000 in l | Path length (cuvette width) | cm | 1 | 0.1 | 10 in mw | Molecular weight (optional) | g/mol | 50000 | 0 | 1000000 out c | Molar concentration | mol/L | a / (e * l) | 6 out cm | Concentration in mg/mL | mg/mL | (a / (e * l)) * mw | 4 out cmm | Concentration in µg/mL | µg/mL | ((a / (e * l)) * mw) * 1000 | 2
Each line is one input (in, sel, list) or one result (out: label, unit, formula, decimals). How the calculators work
Method: the standard formula for this calculation with the constants and conversion factors written on the page, rounded as each result states. See method and constants, the disclaimer or report an error.
More in this section
- QPP Calculator Canada: How It Works & How to Use It
- PPP Calculator World Bank: How It Works & How to Use It2,000.00
- PPP Calculator: India vs USA Cost Comparison1,000.00 USD
- PPP Calculator USD to INR: How It Works2,335 INR
- PPP Calculator India vs UK: Real Cost Comparison103.63 GBP
- PPP Calculator India vs Dubai: Real Cost Comparison8,954.58 AED