molecalculator

Moles & Molar Mass

Centrifuge RPM to RCF: Convert Speed to ×g

Calculator No. 01-2232 · Updated · Formula and assumptions shown below

Relative centrifugal force 11,180 ×g
Speed for target RCF
10,000 rpm
Result, rounded to a whole number
Rotor speed
10000 rpm
Set in the form below
Radius to sample
10 cm
Set in the form below
You enter
Rotor speed (rpm), Radius to sample (cm), Target relative centrifugal force (×g)
You get
Relative centrifugal force (×g), Speed for target RCF (rpm)
Rounding
0 dp

For study and lab planning. The result follows from the numbers you enter and the formula shown. Check it against your method before you weigh, mix or dilute anything, follow the safety data sheet and your supervisor, and do not use it for clinical, dosing or medication work.

Convert centrifuge speed in revolutions per minute (rpm) to relative centrifugal force (RCF, or ×g) using the rotor’s radius: RCF = 1.118 × 10⁻⁵ × r × rpm², where r is the radius in centimeters. Enter rpm and the radius to the sample position to calculate ×g; the reverse conversion is also provided.

Formula

The conversion formula is RCF = 1.118 × 10⁻⁵ × r × rpm². RCF is a dimensionless ratio of centrifugal acceleration to standard gravitational acceleration, commonly written as ×g. In this formula, r is the distance in centimeters from the rotor’s axis of rotation to the point of interest, and rpm is the rotational speed in revolutions per minute. The constant incorporates the conversion from rpm and centimeters to acceleration relative to standard gravity.

To calculate the speed for a chosen force, rearrange the equation: rpm = √[RCF / (1.118 × 10⁻⁵ × r)]. The calculator’s first result is RCF from the entered speed and radius; its second is the rpm needed for the entered target RCF and radius. It treats the two calculations independently.

Worked Example

Using the defaults, the rotor speed is 10,000 rpm and the sample radius is 10 cm:

  • RCF = 1.118 × 10⁻⁵ × 10 × 10,000²
  • 10,000² = 100,000,000
  • RCF = 1.118 × 10⁻⁵ × 10 × 100,000,000 = 11,180 ×g

The reverse calculation uses the default target of 11,180 ×g and radius of 10 cm:

  • rpm = √[11,180 / (1.118 × 10⁻⁵ × 10)]
  • 1.118 × 10⁻⁵ × 10 = 0.0001118
  • rpm = √(100,000,000) = 10,000 rpm

Both outputs are displayed to the nearest whole number, matching the calculator’s settings.

Assumptions

  • The radius is entered in centimeters and is measured from the rotation axis to the sample location being considered.
  • rpm is the rotor’s actual rotational speed, not a percentage of a speed setting or a motor frequency.
  • The standard conversion constant uses standard gravity; RCF is reported as a relative value, not in m/s².
  • The calculator does not account for rotor geometry, tube angle, acceleration or braking profiles, or sample properties.

Why RPM and RCF Are Different

RPM describes how fast something rotates. It does not, by itself, tell you the acceleration experienced by a sample. At the same rpm, a point farther from the axis travels in a larger circle and experiences greater centrifugal acceleration. That is why radius is essential when converting rpm to RCF.

RCF expresses that acceleration as a multiple of standard gravity. A setting of 1,000 ×g means an acceleration magnitude 1,000 times standard gravity at the stated sample position; it does not mean the rotor is turning 1,000 times per minute. The ×g notation is convenient for comparing protocols across rotors of different sizes.

For a fixed rotor radius, doubling rpm makes RCF four times greater because speed is squared in the equation. Doubling radius at the same rpm doubles RCF. These relationships help explain why a seemingly small speed change can substantially change the force applied.

Which Radius Should You Use?

Use the radius specified by the centrifuge or rotor documentation, or by the method you are following. “Radius” is not always a single obvious measurement: the sample occupies a volume, and its position changes with tube shape and rotor design. Common reference points include:

  • Maximum radius: distance from the axis to the bottom or outermost point of the sample. A stated maximum RCF often refers to this point.
  • Average radius: distance to an approximate midpoint of the sample volume; it may be used when a method specifies average RCF.
  • Minimum radius: distance to the top or innermost part of the sample, depending on rotor orientation.

Do not substitute one radius convention for another without checking the protocol. In a fixed-angle rotor, the tube lies at an angle; in a swinging-bucket rotor, it changes orientation during rotation. The relevant position and manufacturer’s definition can therefore differ. Use the same convention as the published method or rotor specification when reproducing a force.

How to Use the Conversion

  1. Find the rotor radius in the appropriate documentation and note whether it is a maximum, average, or other specified radius.
  2. Enter the radius in centimeters. Convert millimeters to centimeters by dividing by 10, or inches to centimeters by multiplying by 2.54.
  3. For a speed-to-force conversion, enter rpm and read the first result, RCF in ×g.
  4. For a force-to-speed conversion, enter the desired ×g and radius, then read the second result, rpm. Confirm that this speed is within the rotor and centrifuge’s permitted limits.

For example, 25 mm is 2.5 cm, not 25 cm. A radius entered in the wrong unit creates a proportional error in the calculated RCF. Likewise, the reverse calculation needs the radius associated with the target force; there is no single rpm corresponding to a target ×g for every rotor.

Common Errors and Safety Limits

A frequent mistake is treating an rpm value as if it were an RCF value, or copying rpm from a method that actually specifies ×g. Another is using the rotor’s outside diameter as its radius. If you have a diameter measurement, divide it by two only when that diameter runs through the axis and the exact sample position of interest.

Calculated values do not establish that a rotor can safely run at that speed. Rotor and centrifuge limits, compatible tubes and adapters, sample balance, loading rules, and rotor condition all matter. Never exceed the rated maximum speed or force, and do not assume that a physically possible calculation is an approved operating setting. Follow the applicable laboratory method, safety data sheet (SDS), and supervisor’s instructions.

For bench work, follow your lab method, the safety data sheet (SDS) and your supervisor. Inspect equipment as required by local procedures, secure the load correctly, and use required containment and protective measures. This arithmetic is for planning and interpretation; it is not a substitute for the instrument’s operating instructions or a validated protocol.

Frequently Asked Questions

How do I convert rpm to RCF?
Multiply the radius in centimeters by rpm squared, then multiply by 1.118 × 10⁻⁵: RCF = 1.118 × 10⁻⁵ × r × rpm².

How do I convert RCF to rpm?
Rearrange the formula: rpm = √[RCF / (1.118 × 10⁻⁵ × r)]. Use the radius in centimeters and the radius convention specified by your method.

Can I convert rpm to ×g without knowing the rotor radius?
No. The same rpm produces different RCF values at different radii, so you need the distance from the rotation axis to the relevant sample position.

Is 1,000 ×g the same as 1,000 rpm?
No. ×g is relative centrifugal force, while rpm is rotational speed. The conversion depends on radius, and RCF also increases with the square of rpm.

Should I use maximum or average radius?
Use whichever radius convention the protocol or rotor specification gives. Maximum and average radii produce different RCF values; do not silently interchange them.

Related

Formula used
in rpm | Rotor speed | rpm | 10000 | 0
in r | Radius to sample | cm | 10 | 0
in g | Target relative centrifugal force | ×g | 11180 | 0
out rcf | Relative centrifugal force | ×g | 1.118e-5 * r * rpm^2 | 0
out speed | Speed for target RCF | rpm | sqrt(g / (1.118e-5 * r)) | 0

Each line is one input (in, sel, list) or one result (out: label, unit, formula, decimals). How the calculators work

Method: the standard formula for this calculation with the constants and conversion factors written on the page, rounded as each result states. See method and constants, the disclaimer or report an error.

More in this section

The whole Moles & Molar Mass section →