Absorbance Equation: A = log10(I0/I) Explained

Calculator No. 01-10478 · Updated · Formula and assumptions shown below

Absorbance (A) 0.4498
Transmittance (T = I/I0)
0.3550
= Transmitted light intensity (I) ÷ Incident light intensity (I0)
Percent transmittance (%T)
35.50 %
= Transmittance (T = I÷I0) × 100
Incident light intensity (I0)
100
Set in the form below
You enter
Incident light intensity (I0), Transmitted light intensity (I)
You get
Absorbance (A), Transmittance (T = I/I0), Percent transmittance (%T) (%)
Rounding
4 dp · 2 dp

For study and lab planning. The result follows from the numbers you enter and the formula shown. Check it against your method before you weigh, mix or dilute anything, follow the safety data sheet and your supervisor, and do not use it for clinical, dosing or medication work.

This page explains the absorbance equation, the core of UV-Vis spectrophotometry. You enter the intensity of incident light (I₀) and transmitted light (I), and the calculator returns absorbance (A), a unitless quantity that is directly proportional to concentration under the Beer-Lambert law. The result tells you how much light a sample absorbs at a given wavelength, which is the basis for quantitative chemical analysis.

Formula

The absorbance equation is:

A = log10(I0 / I)

where:

  • A = absorbance (unitless, often called optical density, OD)
  • I0 = intensity of light incident on the sample (before it enters the sample)
  • I = intensity of light transmitted through the sample (after it exits)

This equation defines absorbance purely from light intensities. It is the logarithmic form of transmittance: T = I / I0, so A = -log10(T). A common alternative form is A = log10(100 / %T), since %T = T × 100.

Worked Example

Using the calculator defaults: I0 = 100 (arbitrary units), I = 35.5.

Step 1: Compute the ratio I0 / I = 100 / 35.5 = 2.8169.

Step 2: Take the base-10 logarithm: log10(2.8169) = 0.4498.

Step 3: The absorbance is 0.4498, which the calculator rounds to 0.4498 (4 decimal places).

Transmittance: T = 35.5 / 100 = 0.355, or 35.5% T. Check: A = -log10(0.355) = 0.4498. The two forms agree.

Assumptions

  • The light is monochromatic (single wavelength).
  • The sample is homogeneous and does not scatter light significantly.
  • The cuvette path length is constant (usually 1 cm).
  • Absorbance is additive for multiple absorbing species at the same wavelength.
  • The detector measures only transmitted light, not reflected or scattered light.

Why the Logarithm? The Beer-Lambert Law

The absorbance equation alone only relates two light intensities. The reason it is so useful is the Beer-Lambert law, which connects absorbance to concentration:

A = ε × c × l

where:

  • ε = molar absorptivity (L·mol−1·cm−1), a substance- and wavelength-specific constant
  • c = molar concentration (mol/L)
  • l = path length of the cuvette (cm, typically 1 cm)

The logarithm is there because each layer of solution absorbs the same fraction of light, not the same amount. If a 1 cm path absorbs 50% of light, a second 1 cm layer absorbs 50% of the remaining light, giving 25% transmitted. This exponential decay becomes a linear relationship when you take the log, which is why absorbance is linear with concentration while transmittance is not. That linearity is what makes calibration curves (A vs. concentration) straight lines.

How to Read the Result

Absorbance values typically range from 0 (no absorption, all light passes through) to values above 3 for very dark samples. In practice:

  • A = 0: I = I₀, no light absorbed.
  • A = 1: 10% of light is transmitted (I = 0.1 × I₀).
  • A = 2: 1% transmitted.
  • A = 3: 0.1% transmitted.

Most spectrophotometers give reliable readings between A = 0.1 and A = 2.5. Below 0.1, the signal is close to the blank and noise dominates. Above 2.5, very little light reaches the detector, so the instrument’s stray light and detector noise cause large errors. If your sample reads above ~2, dilute it and multiply the result by the dilution factor.

Common Mistakes and Unit Pitfalls

  • Using %T directly in the Beer-Lambert law: You must convert %T to absorbance first. A = log10(100 / %T). For example, 50%T gives A = log10(2) = 0.301, not 0.5.
  • Mixing up I₀ and I: I₀ is the blank (solvent-only) reading, I is the sample reading. Swapping them gives a negative absorbance, which is physically meaningless unless the sample fluoresces or the blank was wrong.
  • Forgetting the path length: The Beer-Lambert law uses path length in cm. If your cuvette is 2 cm, the absorbance doubles for the same concentration. Always check the cuvette specification.
  • Using the wrong wavelength: ε varies with wavelength. Absorbance at 280 nm for protein is not the same as at 260 nm for nucleic acids. Always measure at the λmax for your analyte.
  • Ignoring dilution: If you dilute a sample 10-fold before measuring, the measured A corresponds to the diluted concentration. Multiply by the dilution factor to get the original concentration.

Limits of the Equation

The absorbance equation itself is exact for any sample that only absorbs light. The Beer-Lambert law, however, has real limits:

  • High concentration: At concentrations above ~0.01 M, intermolecular interactions and changes in refractive index cause deviations from linearity. The law works best for dilute solutions (A < 1).
  • Stray light: Real instruments have stray light that reaches the detector without passing through the sample, making absorbance read lower than true at high A.
  • Scattering: Turbid samples scatter light, which is counted as absorption. This is why you need a clear solution or a scattering correction.
  • Fluorescence: If the sample fluoresces, the detector may pick up emitted light, reducing apparent absorbance.
  • Chemical equilibria: If the absorbing species participates in a pH-dependent equilibrium (e.g., an indicator dye), the measured absorbance depends on pH, not just total concentration.

For quantitative work, always prepare a calibration curve with known standards under identical conditions (same solvent, pH, temperature, and wavelength), and verify that your sample’s absorbance falls within the linear range of that curve.

Practical Usage in the Lab

In a typical UV-Vis measurement, you first zero the instrument with a blank (the solvent and everything except the analyte). This blank reading is I₀. Then you replace the blank with your sample and read I. The instrument usually displays absorbance directly, but the underlying calculation is exactly the equation above.

Common applications include:

  • Protein quantification: Measure A at 280 nm, use ε or a standard curve (e.g., Bradford or BCA assay).
  • Nucleic acid quantification: A at 260 nm; an A of 1.0 corresponds to ~50 µg/mL double-stranded DNA, ~40 µg/mL RNA, ~33 µg/mL single-stranded DNA (these are approximate, sequence-dependent).
  • Purity checks: The A260/A280 ratio for DNA should be ~1.8; for RNA ~2.0.
  • Kinetics: Measure absorbance change over time to determine reaction rates (e.g., NADH at 340 nm).
  • Colorimetric assays: Many assays produce a colored product whose absorbance is proportional to analyte concentration (e.g., Folin–Ciocalteu for phenolics).

Always follow your lab method, the safety data sheet (SDS) and your supervisor. The absorbance equation gives you the number, but the quality of that number depends on proper blanking, clean cuvettes, and correct wavelength selection.

Frequently Asked Questions

What is the difference between absorbance and optical density?

They are numerically identical: OD = log10(I₀/I). The term “optical density” is often used in microbiology (e.g., OD600 for cell growth) and in older literature. Some instruments display “AU” (absorbance units) or “OD”. The equation and the values are the same.

Can absorbance be greater than 1?

Yes. A = 1 means 10% transmittance; A = 2 means 1% transmittance. There is no upper limit mathematically, but instruments become inaccurate above about A = 2.5 because too little light reaches the detector. For accurate work, keep A below 2 by diluting the sample.

Why is absorbance unitless?

Because it is the logarithm of a ratio of two light intensities (I₀/I), the units cancel. The Beer-Lambert law then gives absorbance the units of concentration × path length × molar absorptivity, but the final A value itself has no units. Some instruments print “AU” or “OD” as a label, not a physical unit.

How do I convert from transmittance to absorbance?

If you have transmittance as a fraction (T), use A = -log10(T). If you have percent transmittance (%T), use A = log10(100 / %T). For example, 50%T → A = log10(2) = 0.301. The calculator above does this automatically.

Why does my absorbance reading change with wavelength?

Because the molar absorptivity ε is a function of wavelength. Each molecule has a unique absorption spectrum with peaks at specific wavelengths. You should measure at the wavelength where ε is maximum (λmax) for the highest sensitivity and the best signal-to-noise ratio.

What is a blank and why do I need it?

A blank is a sample containing everything except the analyte—the solvent, buffer, and any reagents that do not absorb. It sets I₀, so the absorbance you measure for the sample is due only to the analyte. Without a proper blank, your absorbance includes contributions from the solvent and cuvette, giving falsely high values.

Related

Formula used
in i0 | Incident light intensity (I0) | - | 100 | 0.0001
in i | Transmitted light intensity (I) | - | 35.5 | 0.0001
out a | Absorbance (A) | - | log10(i0 / i) | 4
out t | Transmittance (T = I/I0) | - | i / i0 | 4
out p | Percent transmittance (%T) | % | t * 100 | 2

Each line is one input (in, sel, list) or one result (out: label, unit, formula, decimals). How the calculators work

Method: the standard formula for this calculation with the constants and conversion factors written on the page, rounded as each result states. See method and constants, the disclaimer or report an error.