molecalculator

Moles & Molar Mass

Absorbance Unit of Measurement Explained

Calculator No. 01-10375 · Updated · Formula and assumptions shown below

The calculator on this page is switched off while it is checked. The formula is below, and the worked example shows how to do the sum by hand.

For study and lab planning. The result follows from the numbers you enter and the formula shown. Check it against your method before you weigh, mix or dilute anything, follow the safety data sheet and your supervisor, and do not use it for clinical, dosing or medication work.

The absorbance unit of measurement is the absorbance unit (AU), also commonly written as AU or simply left unitless. Absorbance itself is a dimensionless quantity, meaning it has no physical units like grams or meters; the “AU” label is used to clarify that the value comes from a spectrophotometer reading.

Formula

The relationship between absorbance and concentration is given by the Beer-Lambert law:

A = ε × l × c

  • A = absorbance (AU, dimensionless)
  • ε = molar absorptivity (L·mol⁻¹·cm⁻¹), also called the extinction coefficient
  • l = path length of the cuvette (cm, typically 1 cm)
  • c = concentration of the absorbing species (mol/L)

Absorbance is defined as the logarithm (base 10) of the ratio of incident light intensity to transmitted light intensity: A = log₁₀(I₀/I). This means absorbance is a pure number. The AU label is a convenience to distinguish it from transmittance (%), which is also dimensionless but expressed as a percentage.

Worked Example

Using the default values from the calculator: absorbance = 0.5 AU, path length = 1 cm, concentration = 0.0001 mol/L (100 µM).

  1. Rearrange the Beer-Lambert law to solve for ε: ε = A / (l × c)
  2. Substitute the values: ε = 0.5 / (1 × 0.0001) = 0.5 / 0.0001
  3. Calculate: ε = 5000 L·mol⁻¹·cm⁻¹

The calculator also shows transmittance: T = 10⁻ᴬ × 100 = 10⁻⁰·⁵ × 100 ≈ 31.62%. This matches the common expectation that an absorbance of 0.5 corresponds to about 31.6% of light passing through.

Assumptions

  • The Beer-Lambert law assumes a dilute solution (typically < 0.01 M) where molecules do not interact.
  • The light is monochromatic (single wavelength) and the sample is homogeneous.
  • The absorbing species does not scatter light significantly.
  • The path length is constant and known (usually 1 cm).

Why Absorbance Is Unitless (But Labeled AU)

Absorbance is derived from a ratio of light intensities, so the units cancel out. However, spectrophotometer software and scientific papers often append “AU” to remind readers that the number is an absorbance reading, not a concentration or raw intensity. Some instruments display “mAU” (milli-absorbance units), where 1 AU = 1000 mAU. For example, a protein purification chromatogram might show a peak at 250 mAU, which is 0.25 AU.

It is crucial to distinguish absorbance from transmittance. Transmittance (T) is the fraction of light that passes through, often expressed as a percentage. Absorbance is the negative log of transmittance (A = -log₁₀(T)). A transmittance of 100% means no light absorbed (A = 0), while 1% transmittance corresponds to A = 2.

How to Read and Use Absorbance Values

In practice, you measure absorbance with a spectrophotometer at a specific wavelength. The value itself tells you how much light was absorbed, but it is only useful when you know the extinction coefficient (ε) for your molecule at that wavelength. For example, DNA has an extinction coefficient of about 50 (µg/mL)⁻¹ cm⁻¹ at 260 nm, so an absorbance of 1.0 at 260 nm corresponds to roughly 50 µg/mL of double-stranded DNA. For proteins, a common method is the Bradford assay, where absorbance at 595 nm is compared to a standard curve.

To convert absorbance to concentration, use the rearranged Beer-Lambert law: c = A / (ε × l). This is what the calculator does when you input A, ε, and l. Always check the units: if ε is in L·mol⁻¹·cm⁻¹ and l is in cm, then c comes out in mol/L.

Common Mistakes and Misconceptions

  • Confusing absorbance with optical density (OD): OD is often used interchangeably with absorbance, but OD can include scattering losses. In most lab contexts, they are the same.
  • Using the wrong path length: If you use a cuvette with a 2 cm path length, the absorbance doubles for the same solution. Always verify the cuvette’s path length.
  • Forgetting to blank the instrument: You must zero the spectrophotometer with a blank (solvent without sample) to subtract background absorbance.
  • Ignoring the linear range: The Beer-Lambert law is only linear at low concentrations. Above ~1 AU, the relationship becomes nonlinear due to stray light and molecular interactions. If your absorbance is above 1.5, dilute the sample.
  • Mixing up molar absorptivity units: Some sources give ε in mL·mg⁻¹·cm⁻¹ for proteins or nucleic acids. Always convert to L·mol⁻¹·cm⁻¹ if you are using molar concentration.

How to Choose the Right Wavelength and Cuvette

The choice of wavelength depends on the absorption spectrum of your compound. For nucleic acids, use 260 nm; for proteins, 280 nm (aromatic residues) or 595 nm (Bradford assay). For colorimetric assays, the wavelength is usually the complement of the color produced. For example, the Lowry assay measures at 750 nm.

The cuvette material matters for UV measurements: standard glass absorbs UV light below ~350 nm, so you must use quartz cuvettes for wavelengths below that. Plastic cuvettes are fine for visible light (400-700 nm) but may not be UV-transparent. Always check the cuvette’s specifications.

Limitations of Absorbance Measurements

Absorbance is a powerful tool but has limits. It measures total absorption, not the identity of the absorbing species. If your sample contains multiple compounds that absorb at the same wavelength, you get a combined reading. Also, temperature and pH can affect the extinction coefficient. For accurate concentration determination, you need a pure sample and a known ε. The Beer-Lambert law assumes no fluorescence, scattering, or chemical equilibria that shift with concentration.

For turbid samples (e.g., cell suspensions), absorbance readings include scattering, which is why OD600 is used for cell growth but not for absolute concentration. In such cases, the reading is a relative measure of cell density, not a true absorbance.

Frequently Asked Questions

What does 1 AU mean? 1 AU means the sample absorbs 90% of the incident light at that wavelength (since A = 1 corresponds to T = 10%). It is a dimensionless number, but the AU label indicates it is an absorbance reading.

Can absorbance be negative? Yes, if the sample transmits more light than the blank (e.g., the blank absorbs more than the sample). This usually indicates a problem with the blank or the sample, but it can happen if the blank is colored or the sample is diluted.

How do I convert absorbance to percent transmittance? Use the formula T% = 10⁻ᴬ × 100. For A = 0.3, T% = 10⁻⁰·³ × 100 ≈ 50.1%.

What is the difference between absorbance and optical density? In practice, they are often used interchangeably, but OD can include scattering, while absorbance is strictly the log of the ratio of incident to transmitted light. For clear solutions, they are identical.

Why is absorbance measured at 260 nm for DNA? Nucleic acids absorb UV light maximally at 260 nm due to the aromatic bases. The extinction coefficient at that wavelength is well-known, allowing concentration determination.

Is there an upper limit to absorbance? Most spectrophotometers have a linear range up to about 2-3 AU, but the Beer-Lambert law deviates from linearity above ~1 AU due to stray light and instrumental limitations. Dilute your sample to stay in the linear range.

Safety and Best Practices

When measuring absorbance, always wear appropriate personal protective equipment (gloves, safety glasses) if handling chemicals. Follow your lab method, the safety data sheet (SDS) and your supervisor. Ensure the cuvette is clean and free of fingerprints, as these can scatter light and affect readings. Use lint-free wipes for cleaning. For UV measurements, avoid touching the optical surfaces.

Related

Formula used
in a | Absorbance (AU) | - | 0.5 | 0
in l | Path length (cm) | cm | 1 | 0.01
in c | Concentration (M) | mol/L | 0.0001 | 0
out e | Molar absorptivity (L mol⁻¹ cm⁻¹) | L mol⁻¹ cm⁻¹ | a / (l * c) | 0
out t | Transmittance | % | 10^(-a) * 100 | 2

Each line is one input (in, sel, list) or one result (out: label, unit, formula, decimals). How the calculators work

Method: the standard formula for this calculation with the constants and conversion factors written on the page, rounded as each result states. See method and constants, the disclaimer or report an error.

More in this section

The whole Moles & Molar Mass section →