Calculate relative centrifugal force (RCF) from revolutions per minute (rpm) with RCF = 1.118 × 10−5 × r × rpm2, where r is the rotor radius in centimeters. Enter the speed and radius to get RCF in multiples of Earth’s gravity (×g); the result is specific to that radius.
Formula
The standard relationship is:
RCF = 1.118 × 10−5 × r × rpm2
- RCF is relative centrifugal force, expressed as a multiple of standard gravitational acceleration, written ×g. It is a ratio, not a force in newtons.
- r is the distance in centimeters from the axis of rotation to the point in the sample whose RCF you want.
- rpm is rotational speed in revolutions per minute.
- 1.118 × 10−5 is the conversion constant for this particular form of the equation, with radius in centimeters and speed in rpm.
The calculator uses those units directly. If your radius is in millimeters, divide it by 10 to convert to centimeters; if it is in meters, multiply it by 100. Do not enter a radius in centimeters while labeling it as millimeters, because the calculated force would be off by a factor of 10.
Worked Example
Using the calculator’s default values, the rotor speed is 10,000 rpm and the radius is 10 cm.
RCF = 1.118 × 10−5 × 10 × 10,0002
10,0002 = 100,000,000
RCF = 1.118 × 10−5 × 10 × 100,000,000 = 11,180 ×g
The calculator rounds to a whole number, so the displayed result is 11,180 ×g.
Assumptions
- The entered radius is the distance from the rotation axis to the relevant point in the sample, in centimeters.
- The entered rpm is the rotor’s actual rotational speed, not a speed setting expressed in another unit.
- The rotor is rotating steadily, and the reported value is the ideal kinematic RCF at the specified radius.
- The calculator does not account for rotor geometry, sample movement, acceleration or braking, or differences in sedimentation behavior between materials.
What RCF Means and Why Radius Matters
RCF describes the acceleration associated with rotation relative to standard gravity. A value of 1,000 ×g means an acceleration magnitude 1,000 times standard gravitational acceleration at the stated radius. It does not mean that the rotor turns 1,000 times faster, nor is it a direct measure of how quickly a particular particle will settle. Particle size, density, shape, liquid viscosity, and run duration also affect separation.
Radius matters because different positions in a rotor travel in circles of different sizes. At the same rpm, a point farther from the axis has a higher tangential speed and experiences greater centripetal acceleration. Therefore, RCF is not one universal value for an entire rotor: it varies from the center toward the outside. When a protocol specifies an RCF, check which radius convention it uses.
Common conventions include the radius from the axis to the bottom of the sample tube, the midpoint of the sample, or the maximum sample position. These may be called maximum, average, or minimum RCF, depending on the rotor and documentation. Use the radius that matches the specified convention rather than choosing whichever value is easiest to find.
Converting a Protocol Between RPM and RCF
Protocols often specify RCF because it describes the acceleration at a stated radius more directly than rpm does. Rotor models can have different radii, so using the same rpm on two rotors does not necessarily produce the same RCF. If a procedure gives an RCF and you need the corresponding speed, rearrange the equation:
rpm = √(RCF ÷ (1.118 × 10−5 × r))
For example, to find the speed for 5,000 ×g at a radius of 8.0 cm:
rpm = √(5,000 ÷ (1.118 × 10−5 × 8.0)) ≈ 7,474 rpm
This is a mathematical conversion, not a guarantee that a rotor or centrifuge can safely reach that speed. Check the instrument and rotor documentation and follow the specified operating limits. When bench work is involved, follow your lab method, the safety data sheet (SDS), and your supervisor.
Common Errors and Limits
- Using the wrong radius: Measure from the axis of rotation, not from the rotor’s outer edge unless that edge is exactly the sample position required by the protocol.
- Mixing units: This equation requires centimeters and rpm. Convert millimeters or meters before entering the radius.
- Confusing rpm with RCF: RPM is rotational speed; RCF is acceleration relative to gravity. They are related by the square of rpm and by radius.
- Rounding too early: Retain the radius and speed at appropriate precision until the final result. Because speed is squared, a small speed change can noticeably change RCF.
- Assuming equal RCF means identical separation: RCF alone does not account for run time, sample properties, rotor angle, or how a protocol defines its radius.
The formula calculates ideal acceleration at a point. It does not determine a suitable centrifugation condition or validate a method. Use the rotor-specific conversion information where available, especially when translating a published protocol or comparing equipment.
Frequently Asked Questions
How do I calculate RCF from rpm?
Multiply the radius in centimeters by rpm squared, then multiply that product by 1.118 × 10−5. The answer is RCF in ×g.
What radius should I use for RCF?
Use the distance from the rotor’s axis to the sample position specified by the protocol or rotor documentation. The tube bottom, sample midpoint, and outermost sample position can give different values.
Can I convert RCF back to rpm?
Yes. Use rpm = √(RCF ÷ (1.118 × 10−5 × radius in cm)). Confirm that the resulting speed is allowed for the rotor and centrifuge.
Does the same rpm give the same RCF in every rotor?
No. At a fixed rpm, a rotor with a larger radius produces a higher RCF at the corresponding sample position.
Is RCF the same as ×g?
RCF is commonly reported as a multiple of gravity, written ×g. For example, 2,000 ×g means the calculated acceleration is 2,000 times standard gravitational acceleration at the stated radius.
Related
Formula used
in rpm | Rotor speed | rpm | 10000 | 0 in r | Rotor radius | cm | 10 | 0 out rcf | Relative centrifugal force | ×g | 1.118 * 10^-5 * r * rpm^2 | 0
Each line is one input (in, sel, list) or one result (out: label, unit, formula, decimals). How the calculators work
Method: the standard formula for this calculation with the constants and conversion factors written on the page, rounded as each result states. See method and constants, the disclaimer or report an error.
More in this section
- MD Calc I:T Ratio Calculator and Explanation0.0500 ratio
- Oxygen Mass Number: Isotope Values Explained16 nucleons
- Mass-to-Light Ratio: Meaning and Calculation1.000 M☉/L☉
- M/z Ratio in Mass Spectrometry: Meaning and Calculation1,000.00 Th
- Mass Fraction to Molar Fraction Calculator
- K2 Molar Mass Calculator and Calculation78.1966 g/mol