This guide shows you how to convert relative centrifugal force (RCF, often called g-force) to revolutions per minute (RPM) for a centrifuge. You’ll need the g-force value and the rotor’s radius in centimeters. The calculator below does the conversion instantly, and the steps explain the math so you can do it by hand if needed.
Formula
The standard conversion formula is:
RPM = √(RCF × 1000 / (1.118 × 10-5 × r))
Where:
- RCF is the relative centrifugal force in g (unitless, but expressed as “times gravity”).
- r is the rotor radius in centimeters (cm), measured from the center of the rotor to the sample midpoint in the tube.
- RPM is the revolutions per minute.
- 1.118 × 10-5 is a constant derived from the conversion of RPM to radians per second and g to m/s². It assumes the radius is in cm.
This formula comes from the relationship: RCF = 1.118 × 10-5 × r × (RPM)². Rearranging gives the formula above.
Worked Example
Using the default values from the calculator: RCF = 1000 g, radius = 10 cm.
Step 1: Write the formula.
RPM = √(RCF × 1000 / (1.118 × 10-5 × r))
Step 2: Substitute the values.
RPM = √(1000 × 1000 / (1.118 × 10-5 × 10))
Step 3: Calculate the denominator.
1.118 × 10-5 × 10 = 1.118 × 10-4
Step 4: Divide the numerator by the denominator.
1000 × 1000 = 1,000,000
1,000,000 / (1.118 × 10-4) = 8.9445 × 109
Step 5: Take the square root.
√(8.9445 × 109) ≈ 9457.5
So, the rotor speed is approximately 9457.5 RPM. The calculator shows 9457.5 RPM (rounded to one decimal).
Assumptions
- The radius is measured in centimeters from the centrifuge axis to the sample midpoint.
- The g-force is the relative centrifugal force (RCF), not the gravitational constant (9.81 m/s²).
- The conversion uses the standard formula for a fixed-angle or swinging-bucket rotor; the actual g-force at different points in the tube varies slightly, but the midpoint is the convention.
When to Use This Conversion
You need this conversion when a protocol specifies a g-force (e.g., “centrifuge at 10,000 g”) but your centrifuge only displays RPM. Since the required g-force depends on the rotor radius, you must know your rotor’s radius to set the correct RPM. Conversely, if your centrifuge displays g-force, you can use the reverse formula to find the corresponding RPM for a different rotor.
How to Read the Result
The result is the exact RPM setting that produces the desired g-force at the given radius. If your centrifuge has a digital display, set it to this RPM. If your centrifuge has a dial, round to the nearest mark but note that the actual g-force will differ slightly. Always check your centrifuge’s manual for the rotor radius, as using the wrong radius can lead to incorrect g-forces, which may affect sample separation or damage the rotor.
Common Mistakes
- Using the wrong radius: The radius is not the rotor’s diameter or the tube length. It’s the distance from the center of the rotor to the point where the sample sits (usually the middle of the tube).
- Mixing units: The formula requires radius in centimeters. If you measure in inches or millimeters, convert first (1 inch = 2.54 cm, 1 mm = 0.1 cm).
- Forgetting the square root: The formula is not linear. Doubling the g-force does not double the RPM; it increases by a factor of √2.
- Using g-force as grams: g-force is a ratio, not a mass. Do not confuse it with the weight of the sample.
- Ignoring the rotor type: The formula assumes the radius is the same for all tubes, but in a fixed-angle rotor, the radius varies along the tube. Using the midpoint is standard, but for precise work, use the manufacturer’s recommended radius.
Tips
- Always measure the radius with the tube in place, from the center of the rotor to the middle of the liquid column. You can use a ruler or the rotor’s manual.
- If your centrifuge has a g-force setting, use it directly and ignore RPM—it automatically adjusts for the rotor.
- For critical applications (e.g., cell separation), verify the g-force with a tachometer or the centrifuge’s calibration.
- Keep a note of your rotor’s radius on the centrifuge or in your lab notebook so you don’t have to measure each time.
- When in doubt, round up to the nearest RPM that gives a slightly higher g-force, but stay within the rotor’s maximum speed rating.
Troubleshooting
If the result seems too high or too low: Check your radius measurement. A small error in radius can cause a large error in RPM because the formula uses the square root of the radius. Re-measure carefully.
If you get an error from the calculator: Ensure the g-force is positive and the radius is greater than zero. The calculator has minimum limits (0 for g-force, 0.1 cm for radius).
If your centrifuge doesn’t reach the calculated RPM: The rotor may have a maximum speed limit. Check the rotor’s specifications. If the calculated RPM exceeds the limit, you cannot achieve that g-force with that rotor—use a different rotor or reduce the g-force.
If the conversion doesn’t match the centrifuge’s display: Some centrifuges show g-force directly, and the RPM is derived. Use the centrifuge’s own g-force setting if available. Also, confirm the radius you used matches the rotor’s actual radius.
Frequently Asked Questions
What is the difference between g-force and RPM?
RPM is the rotational speed of the rotor (revolutions per minute). G-force (RCF) is the acceleration applied to the sample, which depends on both RPM and the rotor radius. A larger radius creates more g-force at the same RPM.
Why does the radius matter?
The radius determines the distance the sample travels in a circle. The centrifugal force is proportional to the radius times the square of the angular velocity, so a larger radius gives a higher g-force for the same RPM.
Can I convert g-force to RPM without knowing the radius?
No, the radius is essential. Without it, the conversion is impossible because g-force depends on both RPM and radius. You must know the rotor’s radius.
Is the conversion formula the same for all centrifuges?
Yes, the physics is the same, but the radius varies between rotors. Always use the radius specific to your rotor and tube position.
What if my rotor radius is in inches?
Convert inches to centimeters first (1 inch = 2.54 cm) before using the formula. The calculator expects cm.
How accurate is this conversion?
The formula is accurate to the precision of your radius measurement and the centrifuge’s speed control. For most lab purposes, it’s sufficient. For high-precision work, calibrate the centrifuge with a tachometer.
Related
Formula used
in g | G-force (RCF) | g | 1000 | 0 in r | Rotor radius | cm | 10 | 0.1 out rpm | Rotor speed | RPM | sqrt(g * 1000 / (1.118e-5 * r)) | 1
Each line is one input (in, sel, list) or one result (out: label, unit, formula, decimals). How the calculators work
Method: the standard formula for this calculation with the constants and conversion factors written on the page, rounded as each result states. See method and constants, the disclaimer or report an error.
More in this section
- Volume Calculator in Chemistry: How to Use It0.2000 mol/L
- Volume Calculator in mm: Convert & Calculate100,000 mm³
- Volume Calculator Feet and Inches: How to Use It3.12 ft³
- Volume Calculator: Convert ft to Yards1.00 yd³
- Surfboard Volume Calculator JS: How It Works32.4 L
- Volume Calculator cm to Litres: How to Convert2.500 L