The systolic/diastolic (S/D) ratio is a Doppler ultrasound measurement used mainly to assess blood flow resistance in arteries, most commonly the umbilical artery during pregnancy. The normal range depends on the vessel and the gestational age, but for the umbilical artery after 28 weeks, an S/D ratio below 3.0 is generally considered normal, with values decreasing as pregnancy progresses.
Formula
The S/D ratio is the simplest of the Doppler indices. It is calculated by dividing the peak systolic velocity (the highest blood flow velocity during a heartbeat) by the end-diastolic velocity (the lowest velocity just before the next heartbeat):
S/D ratio = Peak systolic velocity / End-diastolic velocity
Where:
- Peak systolic velocity (cm/s) – maximum flow velocity during ventricular contraction.
- End-diastolic velocity (cm/s) – minimum flow velocity just before the next contraction, measured at the end of diastole.
The result is a dimensionless number. A higher ratio means higher downstream resistance to flow. Two related indices are often reported alongside the S/D ratio:
- Resistive index (RI) = (S − D) / S
- Pulsatility index (PI) = (S − D) / mean velocity
These three indices are used because the S/D ratio alone can be misleading when diastolic velocity is very low or zero.
Worked Example
Suppose a Doppler measurement of the umbilical artery gives a peak systolic velocity of 40 cm/s and an end-diastolic velocity of 20 cm/s.
Using the formula:
S/D ratio = 40 / 20 = 2.0
RI = (40 − 20) / 40 = 20 / 40 = 0.50
PI = (40 − 20) / ((40 + 20) / 2) = 20 / 30 = 0.67
These values are within the normal range for a third-trimester pregnancy.
Assumptions
- The measurement is taken with the patient in a stable position (usually semi-recumbent) and during fetal apnea (no fetal breathing movements).
- The Doppler angle of insonation is as close to 0° as possible; angles above 30° introduce significant error.
- The vessel is imaged at a consistent site (e.g., free loop of umbilical cord) and the sample volume covers the entire vessel lumen.
- The values are used for clinical interpretation only in the context of the patient’s full picture – not as a standalone diagnostic test.
Normal Ranges by Gestational Age (Umbilical Artery)
The S/D ratio in the umbilical artery normally falls as the placenta matures and vascular resistance decreases. A commonly cited reference range is:
- 20–28 weeks: 3.0 to 4.5 (upper limit around 4.5)
- 28–32 weeks: 2.8 to 3.5
- 32–36 weeks: 2.5 to 3.0
- 36 weeks to term: 2.0 to 2.8 (values above 3.0 are considered elevated)
These are approximate; each ultrasound unit and reference population may have slightly different nomograms. The key clinical point is the trend: a rising S/D ratio over serial measurements is more concerning than a single borderline value.
Why the S/D Ratio Matters
In obstetrics, the umbilical artery S/D ratio reflects placental vascular resistance. High resistance can indicate placental insufficiency, which may lead to fetal growth restriction, preeclampsia, or stillbirth. When diastolic flow is absent or reversed (a condition called absent end-diastolic flow or reversed end-diastolic flow), the S/D ratio becomes infinite or negative – this is a serious warning sign that requires immediate obstetric review.
The S/D ratio is also measured in other vessels, such as the uterine artery, middle cerebral artery, and renal artery, but normal values differ. For example, in the uterine artery, a high S/D ratio (often above 2.6) after 24 weeks may indicate increased risk of preeclampsia. In the middle cerebral artery, a low S/D ratio (or low PI) can indicate fetal anemia or brain-sparing effect.
How to Choose the Right Doppler Index
While the S/D ratio is easy to compute, it has limitations. When diastolic velocity is very low, the ratio becomes extremely sensitive to small measurement errors. For example, if diastolic velocity is 5 cm/s, a 1 cm/s error changes the ratio from 8.0 to 10.0. In such cases, the pulsatility index (PI) is preferred because it uses the mean velocity, which is more stable.
Clinical guidelines often recommend using the PI or RI instead of the S/D ratio for assessing placental function, especially in high-risk pregnancies. However, the S/D ratio remains widely used because it is intuitive and easy to explain to patients.
Common Mistakes and Pitfalls
- Not correcting for fetal heart rate: The S/D ratio varies with heart rate; a slower heart rate tends to increase the ratio. Some nomograms adjust for this.
- Measuring at the wrong site: The umbilical artery should be sampled at a free-floating loop of cord, not near the placental insertion or fetal abdomen, where values differ.
- Ignoring the Doppler angle: If the angle is >30°, the velocity is underestimated, and the ratio changes. Always keep the angle as low as possible.
- Using a single measurement: A single elevated value is not diagnostic; serial measurements are needed to assess trends.
- Confusing the S/D ratio with the RI or PI: They are related but not interchangeable. The S/D ratio has no upper limit, while RI and PI are bounded (RI between 0 and 1).
Limitations and Clinical Context
The S/D ratio is a screening tool, not a diagnostic test. A normal ratio does not guarantee a healthy pregnancy, and an elevated ratio does not always indicate pathology. Factors such as maternal position, fetal breathing, uterine contractions, and even the specific ultrasound machine can affect the measurement.
In non-obstetric settings, the S/D ratio is used to assess renal artery stenosis, carotid artery disease, and other vascular conditions. Normal values vary by vessel and age. For example, in the renal artery, an S/D ratio above 3.5 may suggest stenosis. Always refer to vessel-specific reference tables.
Finally, the S/D ratio is just one piece of the puzzle. Clinicians combine it with other Doppler indices, biophysical profile, and fetal growth parameters to make decisions. If you are a patient, discuss your results with your healthcare provider – they can interpret them in the context of your entire pregnancy.
Frequently Asked Questions
What does a high S/D ratio mean in pregnancy?
A high S/D ratio (above the 95th percentile for gestational age) suggests increased resistance in the placental circulation. This can happen with preeclampsia, fetal growth restriction, or umbilical cord abnormalities. Your doctor will likely recommend closer monitoring and possibly additional tests.
What is the normal S/D ratio at 30 weeks?
At 30 weeks, the normal S/D ratio for the umbilical artery is typically between 2.5 and 3.5. Values above 3.5 are considered elevated and warrant further evaluation.
Can the S/D ratio be too low?
Yes, an abnormally low S/D ratio (e.g., below 2.0 in the third trimester) can occur with fetal anemia or certain fetal heart arrhythmias. However, low values are less common and less concerning than high values.
Is the S/D ratio the same as the resistive index?
No. The S/D ratio is systolic divided by diastolic velocity, while the resistive index is (S − D)/S. They are related but measure different aspects of the waveform. The RI is always between 0 and 1, while the S/D ratio can be any number above 1.
How is the S/D ratio measured?
It is measured using Doppler ultrasound, which records the velocity of blood flow over time. The machine automatically calculates the peak systolic and end-diastolic velocities, and the ratio is computed by the software or manually.
Does the S/D ratio change with fetal movement?
Yes, fetal breathing movements and fetal heart rate variations can affect the measurement. For this reason, the measurement is typically taken during fetal apnea (no breathing) and with a stable heart rate.
Related
Formula used
in m | Systolic velocity (cm/s) | cm/s | 40 | 0 in d | Diastolic velocity (cm/s) | cm/s | 20 | 0 out sd | S/D ratio | - | m / d | 2 out pi | Pulsatility index (PI) | - | (m - d) / ((m + d) / 2) | 2 out ri | Resistive index (RI) | - | (m - d) / m | 2
Each line is one input (in, sel, list) or one result (out: label, unit, formula, decimals). How the calculators work
Method: the standard formula for this calculation with the constants and conversion factors written on the page, rounded as each result states. See method and constants, the disclaimer or report an error.
More in this section
- Normality to g/L Converter & Formula29.220 g/L
- Normality to meq/mL Conversion Calculator1.0000 meq/mL
- Normality Test Q-Q Plot: How to Read It
- Normality Test Results: How to Analyze Them
- Normality Slay the Spire?
- Normality Test for Paired t Test: How to Check