Absorbance measures how much light a sample blocks, and the SI unit for absorbance is actually none — it is a dimensionless quantity. You enter the intensity of light before and after it passes through a sample, and the calculator returns the absorbance value, which has no physical unit. This page explains why absorbance has no SI unit, how the logarithmic definition works, and how to interpret absorbance values in spectroscopy.
Formula
Absorbance is defined by the equation:
A = log10(I0 / I)
- A – absorbance (dimensionless)
- I0 – intensity of the incident light (same unit as I, typically W/m² or a relative count)
- I – intensity of the transmitted light (same unit as I0)
Because I0 and I share the same unit, they cancel in the ratio, leaving a pure number. The logarithm of a pure number is also a pure number, so absorbance has no SI unit. It is often reported in “absorbance units” (AU), but that is a label, not an SI unit.
Worked Example
Using the default values from the calculator: I0 = 100 units, I = 1 unit.
- Calculate the ratio: I0 / I = 100 / 1 = 100
- Take the base-10 logarithm: log10(100) = 2
- So A = 2.0000
The transmittance is (I / I0) × 100 = (1/100) × 100 = 1.00%. This matches the calculator output.
Assumptions
- The light is monochromatic (single wavelength).
- The sample is homogeneous and does not scatter light significantly.
- The intensity values are measured under the same conditions (same detector, same path length).
- The relationship between absorbance and concentration follows the Beer–Lambert law (linear at low concentrations).
Why Absorbance Has No SI Unit
The International System of Units (SI) defines seven base units: metre, kilogram, second, ampere, kelvin, mole, and candela. Derived units are built from these, such as m/s for speed or kg/m³ for density. Absorbance does not fit into this framework because it is a ratio of two identical quantities.
When you divide two numbers that have the same unit, the unit cancels. For example, if I0 = 5 mW and I = 1 mW, the ratio is 5 (not 5 mW). Taking a logarithm further removes any possibility of a unit, because logarithms are only defined for dimensionless numbers. This is why absorbance is classified as a dimensionless quantity in metrology.
In practice, you will see “AU” (absorbance units) or “OD” (optical density) on spectrophotometers. These are not SI units but rather shorthand to remind you that the number came from a logarithmic scale. The SI brochure itself does not list absorbance as a derived unit; it is simply a quantity of dimension one.
How to Read Absorbance Values
Absorbance is a logarithmic scale, so each unit increase means the sample transmits ten times less light. For example:
- A = 0 means all light passes through (I = I0).
- A = 1 means 10% of light passes through (I = 0.1 × I0).
- A = 2 means 1% passes through.
- A = 3 means 0.1% passes through.
Most spectrophotometers give reliable readings between A = 0.1 and A = 2.5. Below 0.1, the signal-to-noise ratio is poor; above 2.5, stray light and detector nonlinearity cause deviations from the Beer–Lambert law. If your sample gives an absorbance above 2.5, dilute it and multiply the result by the dilution factor.
Common Mistakes and Misconceptions
A frequent error is confusing absorbance with transmittance. Transmittance (T) is the fraction of light that passes through, expressed as a percentage. Absorbance is the negative logarithm of transmittance: A = -log10(T). For T = 1% (0.01), A = -log10(0.01) = 2. People sometimes forget the negative sign or use natural log instead of base-10; the Beer–Lambert law uses base-10, while some physics contexts use natural log.
Another misconception is that absorbance can be greater than 1 and still be accurate. While the number can be any positive value, instrument limitations mean that very high absorbance readings are unreliable. Also, absorbance is additive only when using the same wavelength and path length: Atotal = A1 + A2 for two absorbing species in the same solution, provided they do not interact.
Finally, some sources call absorbance “optical density” (OD). They are numerically identical, but OD is often used in microbiology and biochemistry to describe turbidity, which can include scattering. For clear solutions, absorbance and OD are the same.
Absorbance in the Beer–Lambert Law
The most common use of absorbance is to find concentration via the Beer–Lambert law:
A = ε × c × l
- ε – molar absorptivity (L·mol⁻¹·cm⁻¹)
- c – concentration (mol/L)
- l – path length (cm)
Here, the units of ε, c, and l multiply to give a dimensionless number. This works because ε has units that cancel the units of c and l. For example, if ε = 15000 L·mol⁻¹·cm⁻¹, c = 0.0001 mol/L, and l = 1 cm, then A = 15000 × 0.0001 × 1 = 1.5. The units cancel perfectly.
This is why absorbance is so useful: it is directly proportional to concentration, so a calibration curve of A versus concentration gives a straight line through the origin. You can then measure an unknown sample’s absorbance and read its concentration from the line.
Limits of the Unitless Definition
The dimensionless nature of absorbance is elegant, but it has practical limits. Because it is a ratio, the actual intensity units do not matter — you could use photons per second, watts, or arbitrary counts. However, the instrument must be calibrated to zero with a blank (the solvent without the analyte). If the blank absorbs light, it will affect the readings.
Another limit is that the Beer–Lambert law only holds for dilute solutions (typically below 0.01 mol/L). At high concentrations, intermolecular interactions, changes in refractive index, and aggregation can cause deviations. Also, the law assumes monochromatic light; if the light source has a broad bandwidth, the measured absorbance will be lower than the true value.
Finally, remember that absorbance is wavelength-dependent. The same sample will have different absorbance at different wavelengths, so always specify the wavelength when reporting a value (e.g., “A260 = 0.5″).
Frequently Asked Questions
Is absorbance measured in AU?
Yes, but AU is not an SI unit. It is a label used to remind you that the value is on an absorbance scale. The SI system treats absorbance as dimensionless, so you can write A = 0.5 without any unit. In practice, you will see “0.5 AU” on instrument displays, but it is equivalent to just 0.5.
Why is absorbance unitless while concentration has units?
Because absorbance is a ratio of two light intensities. The units of intensity cancel, leaving a pure number. Concentration is a ratio of amount to volume, but those are different quantities, so the units do not cancel. The Beer–Lambert law connects the two by having molar absorptivity carry the necessary units.
Can absorbance be negative?
Yes, but it is rare. A negative absorbance means the sample transmits more light than the blank. This can happen if the blank absorbs more than the sample, or if there is a fluorescence or scattering artifact. In practice, negative absorbance often indicates a problem with the blank or the instrument.
What is the difference between absorbance and transmittance?
Transmittance is the fraction of light that passes through (I/I0), usually expressed as a percentage. Absorbance is the negative base-10 logarithm of transmittance. They are inversely related: high transmittance means low absorbance, and vice versa. The calculator above shows both values.
Does absorbance have a symbol in the SI?
No, the SI does not assign a special symbol for absorbance. The quantity is simply called “absorbance” and is often denoted by A. The SI brochure lists it as a dimensionless quantity, but it is not a defined derived unit.