This guide shows you how to convert percent transmittance (%T) into absorbance (A), a value used in spectrophotometry to measure how much light a sample absorbs. You’ll learn the simple logarithmic formula, see a worked example, and get tips for accurate results.
Formula
The relationship between transmittance and absorbance is:
A = -log₁₀(T)
where:
- A = absorbance (dimensionless, often reported in absorbance units, AU)
- T = transmittance, expressed as a decimal fraction (e.g., 0.50 for 50%T)
Since %T is just T × 100, the formula becomes:
A = -log₁₀(%T / 100)
For example, if %T = 50, then T = 0.50, and A = -log₁₀(0.50) = 0.3010.
Worked Example
Using the calculator defaults: %T = 50.
- Convert %T to a decimal: T = 50 / 100 = 0.50
- Take the base‑10 logarithm: log₁₀(0.50) = -0.3010
- Apply the negative sign: A = -(-0.3010) = 0.3010
So, 50% transmittance corresponds to an absorbance of 0.3010 AU.
Assumptions
- The measurement is made at a single wavelength.
- The sample is in a standard cuvette with a fixed path length (usually 1 cm).
- The instrument is properly blanked (zeroed) with a reference solution.
- Transmittance is measured as a percentage between 0 and 100.
When to Use This Conversion
Use this conversion whenever you have a %T reading from a spectrophotometer and need the absorbance value for calculations like Beer‑Lambert law (A = εcl) to find concentration. Many instruments display both, but sometimes you only get %T from older machines or from published data.
How to Read the Result
Absorbance values are typically between 0 and 2 for reliable measurements. A value of 0 means no light is absorbed (100%T). A value of 1 means 90% of the light is absorbed (10%T). A value of 2 means 99% absorbed (1%T). Values above 2 are possible but often less accurate because very little light reaches the detector.
Common Mistakes
- Forgetting to divide by 100 – Using %T directly in the log gives a wrong result. Always convert to decimal first.
- Using natural log instead of log₁₀ – Absorbance uses base‑10 logarithm. Using ln will give a different number.
- Misreading the scale – Ensure your %T reading is actually a percentage, not a fraction. If the instrument shows 0.50, that may already be T, not %T.
- Ignoring the blank – If you don’t zero the instrument with a blank, your %T will be off, and the absorbance will be incorrect.
Tips
- Always record the wavelength used, because absorbance is wavelength‑dependent.
- For best accuracy, keep absorbance between 0.1 and 1.5. If your sample is too concentrated, dilute it and account for the dilution factor.
- Use the same cuvette for blank and sample, or ensure matched cuvettes, to avoid path‑length differences.
- When reporting, include the path length (e.g., “A = 0.301 at 450 nm, 1 cm cell”).
Troubleshooting
My absorbance value seems too high or too low. Check that your %T is between 0 and 100. If %T is 100, absorbance is 0. If %T is 0, absorbance is infinite (in practice, very high). Also verify you used base‑10 log, not natural log.
I get a negative absorbance. This can happen if %T > 100 (possible if the blank is absorbing more than the sample). Re‑zero the instrument with a fresh blank.
The result is “Cannot compute”. This occurs if %T is 0 or negative. Transmittance cannot be zero or negative in a valid measurement. Check your input.
My absorbance values are not linear with concentration. This is often due to stray light, high concentrations, or chemical deviations. Dilute the sample and re‑measure.
Frequently Asked Questions
Q: What is the difference between transmittance and absorbance?
A: Transmittance is the fraction of light that passes through a sample, while absorbance is a logarithmic measure of how much light is absorbed. They are inversely related: higher absorbance means lower transmittance.
Q: Can I convert %T to absorbance without a calculator?
A: Yes, you can use a logarithm table or estimate. For common values: 100%T → A=0, 10%T → A=1, 1%T → A=2. For other values, you need a log calculation.
Q: Why does absorbance have no units?
A: Absorbance is defined as a logarithm of a ratio, so it is dimensionless. It is often written as “AU” (absorbance units) for clarity, but it’s not a physical unit.
Q: What is the Beer‑Lambert law?
A: A = εcl, where ε is the molar absorptivity (L·mol⁻¹·cm⁻¹), c is concentration (mol/L), and l is path length (cm). It relates absorbance to concentration.
Q: Does temperature affect absorbance?
A: Yes, temperature can change the sample’s volume and chemical equilibrium, affecting absorbance. For precise work, keep the temperature constant.
Q: What if my instrument gives %T as a fraction like 0.5?
A: If the display shows 0.5, that is likely transmittance (T), not %T. You would use A = -log₁₀(0.5) = 0.3010. If it shows 50, that’s %T, and you divide by 100 first.
Related
Formula used
in p | Percent transmittance (%T) | % | 50 | 0.001 | 100 out a | Absorbance | AU | -log10(p / 100) | 4
Each line is one input (in, sel, list) or one result (out: label, unit, formula, decimals). How the calculators work
Method: the standard formula for this calculation with the constants and conversion factors written on the page, rounded as each result states. See method and constants, the disclaimer or report an error.
More in this section
- Rate and Extent of Reaction: Key Facts0.050 mol/L
- How to Calculate Extent of Reaction0.800 mol
- Rate and Extent of Chemical Reactions0.1000 mol/(L·s)
- Chemical Reaction Formula: How to Write & Balance2.2722 mol
- Extent of Reaction Solver & How to Use It0.6000 mol
- Extent of Reaction Units: mol and Beyond