This calculator converts between relative centrifugal force (RCF), commonly called “g force,” and revolutions per minute (RPM) for a centrifuge rotor. You enter the rotor radius in millimeters and either the desired g force or the RPM, and the calculator returns the corresponding value using the standard centrifuge formula.
Formula
The relationship between RPM and g force is:
RCF (g) = 1.118 × 10-5 × r × (RPM)2
And rearranged to solve for RPM:
RPM = √(RCF / (1.118 × 10-5 × r))
Where:
- RCF = relative centrifugal force, expressed as multiples of Earth’s gravity (g)
- r = rotor radius in millimeters (mm), measured from the center of the rotor to the bottom of the tube when it is in the rotor
- RPM = revolutions per minute of the rotor
The constant 1.118 × 10-5 comes from converting RPM to radians per second and millimeters to meters, then dividing by the standard acceleration due to gravity (9.80665 m/s²).
Worked Example
Using the default values from the calculator: rotor radius r = 100 mm, desired g force = 1000 g.
To find the required RPM:
- Plug the numbers into the rearranged formula: RPM = √(1000 / (1.118 × 10-5 × 100))
- Calculate the denominator: 1.118 × 10-5 × 100 = 1.118 × 10-3
- Divide: 1000 / 1.118 × 10-3 = 894,454.4
- Take the square root: √894,454.4 ≈ 945.8
- So the required rotor speed is approximately 946 RPM.
Checking the reverse: 1.118 × 10-5 × 100 × 946² = 1.118 × 10-5 × 100 × 894,916 = 1,000.4 g, which rounds to 1000 g.
Assumptions
- The rotor radius is measured to the bottom of the tube at its maximum angle (the maximum radius).
- The centrifuge is operating at steady speed, not during acceleration or deceleration.
- The g force is uniform across the sample – in reality, it varies along the tube length; the maximum is at the bottom.
- The formula uses standard gravity (9.80665 m/s²) and ignores any small variations due to altitude or latitude.
Why the Radius Matters
The same RPM produces very different g forces on different rotors. A microcentrifuge rotor with a radius of 50 mm at 10,000 RPM gives 5,590 g, while a large swinging-bucket rotor with a 150 mm radius at the same speed gives 16,770 g. This is why protocols always specify g force, not just RPM. If you see a protocol that says “centrifuge at 12,000 × g for 10 minutes,” you must know your rotor’s radius to convert that to the RPM setting on your machine.
Most centrifuge manuals list the rotor radius for each rotor model. If you don’t have the manual, you can measure it: open the rotor, measure from the center of the rotor spindle to the bottom of the tube hole when the tube is fully inserted. For angle rotors, the bottom of the tube is at the outer edge; for swinging buckets, it’s the bottom of the bucket when horizontal.
How to Use the Calculator
Enter your rotor radius in millimeters. Then either enter a desired g force to get the RPM, or enter an RPM to get the g force. The calculator shows both results simultaneously, so you can see the relationship. For example, with a 100 mm radius:
- 500 RPM → 28 g
- 1,000 RPM → 112 g
- 3,000 RPM → 1,006 g
- 10,000 RPM → 11,180 g
Notice that doubling the RPM quadruples the g force because of the square relationship. This means small changes in speed have large effects on the force applied to your sample.
Common Mistakes
Using the wrong radius unit. The formula requires millimeters. If you measure in centimeters (common on some rotors), multiply by 10 before entering. If you measure in meters, multiply by 1000.
Forgetting the square. Some people accidentally use RPM × radius instead of RPM² × radius. This gives a result that is far too low. Always square the RPM first.
Using average radius instead of maximum. The g force at the top of the tube is lower than at the bottom. Most published protocols refer to the maximum g force (at the bottom). If you use the average radius, you’ll get a lower value that doesn’t match the protocol.
Rounding too early. The constant 1.118 × 10-5 is already rounded. If you use 1.12 × 10-5, you introduce about 0.2% error, which is usually negligible but can matter for precise work. Keep the full constant in intermediate steps.
Limits of the Formula
This formula is a physical relationship that holds for any centrifuge rotor, but it assumes a rigid rotor and a fixed radius. In practice, rotors expand slightly at high speeds due to centrifugal force, which can increase the radius by a few percent. For most laboratory work, this effect is negligible, but for ultracentrifuges operating above 50,000 RPM, manufacturers provide correction charts or software that account for rotor stretch.
Also, the g force experienced by the sample is not uniform. The top of the liquid column experiences less force than the bottom. For pelleting cells or debris, the maximum g force is the relevant value because that’s where the pellet forms. For gradient separations, the g force varies across the gradient, and you may need to calculate the force at a specific position.
Frequently Asked Questions
What is the difference between g force and RPM? RPM is the rotational speed of the rotor (revolutions per minute). G force (RCF) is the acceleration applied to the sample relative to Earth’s gravity. The same RPM can produce different g forces depending on the rotor radius. G force is the physically meaningful value for sedimentation.
How do I convert g to RPM without a calculator? You can use the formula RPM = √(g / (1.118 × 10-5 × r)) where r is in mm. For a quick estimate, you can approximate: for a 100 mm radius, 1,000 g is about 946 RPM, 10,000 g is about 2,991 RPM. But for accuracy, always use the formula with your exact radius.
Why do some protocols say “× g” instead of RPM? Because g force is independent of the centrifuge model. If a protocol specified RPM, it would only be correct for the exact rotor used in that lab. By specifying g force, anyone can adapt the protocol to their own centrifuge by converting using their rotor radius.
Can I use this calculator for any centrifuge? Yes, as long as you know the rotor radius. The formula applies to all centrifuges – microcentrifuges, benchtop, floor models, and ultracentrifuges. Just make sure you use the radius in millimeters and that the rotor is not stretched (which is only significant at very high speeds).
What if my rotor radius is not listed? Measure it. For a fixed-angle rotor, measure from the center of the rotor to the bottom of the tube hole when the tube is inserted. For a swinging-bucket rotor, measure from the center to the bottom of the bucket when it is horizontal. If you can’t measure, contact the manufacturer or check the rotor’s manual – the radius is usually printed on the rotor or in the documentation.
Always follow your laboratory’s standard operating procedures, the centrifuge manufacturer’s instructions, and the safety data sheet (SDS) for the materials you are handling. Never exceed the maximum speed rating of your rotor.
Related
Formula used
in r | Rotor radius | mm | 100 | 0 in g | Desired g force | g | 1000 | 0 in rpm | Rotor speed | RPM | 3000 | 0 out gcalc | Calculated g force | g | 1.118e-5 * r * rpm^2 | 1 out rpmcalc | Required RPM | RPM | sqrt(g / (1.118e-5 * r)) | 0
Each line is one input (in, sel, list) or one result (out: label, unit, formula, decimals). How the calculators work
Method: the standard formula for this calculation with the constants and conversion factors written on the page, rounded as each result states. See method and constants, the disclaimer or report an error.
More in this section
- Motor RPM from Hz: How to Calculate It1,800 RPM
- Rotor Diameter Ratio: What It Means & Why It Matters3.000
- Base Rate Rotor Wing: Meaning & Use25.0 %
- Rotor Diameter: Measurement & Meaning27.49 m/s
- Rotor Diameter Calculator: How to Find It0.1592 m
- Rotor Diameter Formula & How to Calculate It17.00 m